Lesion Durability of the Second Generation Pentaspline Pulsed Field Ablation Catheter <i>NAVIGATE-PF</i> Phase 2 Results
BACKGROUND: Remapping studies have evaluated chronic lesion durability of the first-generation pentaspline pulsed field ablation (PFA) catheter that did not have integrated electroanatomical mapping. By invasive remapping, this first-in-human study evaluated the durability of lesions facilitated by the integration of the navigation-enabled pentaspline PFA catheter and mapping system in patients with atrial fibrillation. METHODS: trial evaluated the second-generation, mapping-integrated pentaspline PFA catheter. Lesion durability was assessed by invasive remapping 60 days post index procedure. RESULTS: After index ablation in 20 patients with the pentaspline PFA catheter, remaps were performed 60 days later in 17 patients, demonstrating a high rate of lesion durability in the pulmonary veins (95%) and posterior wall (86%). CONCLUSIONS: By facilitating real-time visualization of catheter positioning and acute lesion estimation through tags, the second-generation map-and-ablate pentaspline PFA catheter and integrated mapping system may enhance procedural success. REGISTRATION: URL: https://www.clinicaltrials.gov; Unique identifier: NCT06175234.
Authors
- Brynn Okeson (ORCID: https://orcid.org/0000-0002-6912-7259)
- E Mikoś
- Moritoshi Funasako (ORCID: https://orcid.org/0000-0002-1620-3924)
- Tobias Oesterlein (ORCID: https://orcid.org/0000-0002-1495-4927)
- Sarah R. Gutbrod
- Brendan Koop (ORCID: https://orcid.org/0009-0006-5786-627X)
- Jan Petrů
- Camille Metzdorff
- Vivek Y. Reddy
Institutions
- Boston Scientific (United States) (US)
- Mount Sinai Hospital (US)
- Na Homolce Hospital (CZ)
Publication Details
- Journal
- Circulation Arrhythmia and Electrophysiology
- Published
- 2026-04-01
- DOI
- https://doi.org/10.1161/circep.125.014477
- Primary Topic
- Cardiac Arrhythmias and Treatments
- Type
- article
- Field-Weighted Citation Impact
- 0.00