Objective: Despite the universal calibration of commercial IGF1 immunoassays to WHO IS 02/254, substantial inter-assay variability persists, leading to inconsistent patient classification. Harmonization towards a higher-order analytical anchor may reduce such variability. Methods: Four matrix-matched, multi-level serum reference materials (RMs) were prepared from donor serum and value-assigned using an LC-MS/MS method calibrated to WHO IS 02/254. Commutability was assessed according to IFCC recommendations across four immunoassays (Cobas, iSYS, Immulite, Liaison). Deming regression-based recalibration equations derived from commutable RMs were applied to patient samples and healthy donor samples. The primary quantitative endpoint was reduction in standard error of estimate (SEE) relative to the LC-MS/MS method. Age- and sex-specific LC-MS/MS-anchored reference intervals were constructed as a downstream application. Results: Prior to recalibration, immunoassays showed positive bias relative to LC-MS/MS of up to 60%. All four RMs were commutable for Liaison and iSYS, whereas the lowest concentration RM was classified as non-commutable for Cobas and Immulite. Recalibration towards the LC-MS/MS anchor resulted in marked alignment towards the identity line and reduced pooled SEE from 7.82 to 4.89 nmol/L (-37.4%) in patient samples and from 7.34 to 2.09 nmol/L (-71.5%) in healthy samples. Although harmonization effects were assay-dependent at the individual platform level, overall cross-platform dispersion was substantially attenuated. Conclusions: Matrix-matched, commutable serum RMs value-assigned by a higher-order LC-MS/MS procedure enable substantial reduction of inter-assay bias and variability among IGF1 immunoassays. Harmonization towards a higher-order analytical anchor is achievable in routine practice and provides a robust foundation for consistent cross-platform interpretation.

Authors

Institutions

Publication Details

Journal
Endocrine Connections
Published
2026-04-01
DOI
https://doi.org/10.1530/ec-25-0805
Primary Topic
Growth Hormone and Insulin-like Growth Factors
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Harmonization of IGF1 immunoassays towards a higher-order LC-MS/MS reference anchor

Egwm Lentjes, Martijn van Faassen, IP Kema, R.E.A. Musson et al.
Endocrine Connections
Growth Hormone and Insulin-like Growth Factors
article

Harmonization of IGF1 immunoassays towards a higher-order LC-MS/MS reference anchor

Egwm Lentjes, Martijn van Faassen, IP Kema, R.E.A. Musson, M.J.Vos, M.S Pratt
article en

Abstract

Objective: Despite the universal calibration of commercial IGF1 immunoassays to WHO IS 02/254, substantial inter-assay variability persists, leading to inconsistent patient classification. Harmonization towards a higher-order analytical anchor may reduce such variability. Methods: Four matrix-matched, multi-level serum reference materials (RMs) were prepared from donor serum and value-assigned using an LC-MS/MS method calibrated to WHO IS 02/254. Commutability was assessed according to IFCC recommendations across four immunoassays (Cobas, iSYS, Immulite, Liaison). Deming regression-based recalibration equations derived from commutable RMs were applied to patient samples and healthy donor samples. The primary quantitative endpoint was reduction in standard error of estimate (SEE) relative to the LC-MS/MS method. Age- and sex-specific LC-MS/MS-anchored reference intervals were constructed as a downstream application. Results: Prior to recalibration, immunoassays showed positive bias relative to LC-MS/MS of up to 60%. All four RMs were commutable for Liaison and iSYS, whereas the lowest concentration RM was classified as non-commutable for Cobas and Immulite. Recalibration towards the LC-MS/MS anchor resulted in marked alignment towards the identity line and reduced pooled SEE from 7.82 to 4.89 nmol/L (-37.4%) in patient samples and from 7.34 to 2.09 nmol/L (-71.5%) in healthy samples. Although harmonization effects were assay-dependent at the individual platform level, overall cross-platform dispersion was substantially attenuated. Conclusions: Matrix-matched, commutable serum RMs value-assigned by a higher-order LC-MS/MS procedure enable substantial reduction of inter-assay bias and variability among IGF1 immunoassays. Harmonization towards a higher-order analytical anchor is achievable in routine practice and provides a robust foundation for consistent cross-platform interpretation.

Endocrine ConnectionsVol. 15(4)
University Medical Center Groningen (NL), University of Groningen (NL), Utrecht University (NL), University Medical Center Utrecht (NL), Center for Assessment (US)
Openalex Percentile: Top 38%
Growth Hormone and Insulin-like Growth Factors
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.