Detecting Uniformity Artifacts in Ultrasound Transducers

OBJECTIVES: Uniformity artifacts caused by defective transducer elements or scanner malfunctions degrade diagnostic image quality. Traditional quality control (QC) methods, such as phantom testing and visual or quantitative image analysis, can be labor-intensive and limited in test frequency. This study aims to develop a deep learning framework to detect uniformity artifacts and complement traditional QC. METHODS: Clinical median images were generated by aggregating co-registered grayscale ultrasound images acquired by each transducer and computing median intensity values across the image stack. A pretrained ResNet-18 model was fine-tuned on a dataset consisting of clinical median images from linear and curvilinear transducers. The dataset was divided into training, validation, and testing subsets, ensuring no overlap between training and test transducers. To assess generalizability, the model was also evaluated on an independent test set of 396 phantom-validated images from linear and curvilinear transducers. RESULTS: The model achieved 100% accuracy on the first dataset's test set. On the independent test set, it attained 87.4% accuracy with high sensitivity (0.84) and specificity (0.88), demonstrating robust generalization. Occlusion sensitivity maps confirmed the model's attention to uniformity artifact regions. CONCLUSION: The deep learning framework using clinical median images demonstrated robust performance across several linear and curvilinear transducer models. It could be integrated into the clinical QC workflow by automating artifact detection in an effective and timely manner. In our practice, it can flag median images classified as artifact-present, reducing human review time by approximately 80% while preserving detection accuracy.

Authors

Institutions

Publication Details

Journal
Journal of Ultrasound in Medicine
Published
2026-04-01
DOI
https://doi.org/10.1002/jum.70239
Citations
1
Primary Topic
Ultrasound Imaging and Elastography
Type
article
Field-Weighted Citation Impact
24.37
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Detecting Uniformity Artifacts in Ultrasound Transducers

Scott F. Stekel, Donald J. Tradup, Zaiyang Long, Chunming Gu et al.
1 citations
Journal of Ultrasound in Medicine
Ultrasound Imaging and Elastography
24.37
article

Detecting Uniformity Artifacts in Ultrasound Transducers

Scott F. Stekel, Donald J. Tradup, Zaiyang Long, Chunming Gu, Kevin M. Brom, Jaydev Dave, Nicholas J. Hangiandreou, Zhuoyao Xin
article en
1 citations

Abstract

OBJECTIVES: Uniformity artifacts caused by defective transducer elements or scanner malfunctions degrade diagnostic image quality. Traditional quality control (QC) methods, such as phantom testing and visual or quantitative image analysis, can be labor-intensive and limited in test frequency. This study aims to develop a deep learning framework to detect uniformity artifacts and complement traditional QC. METHODS: Clinical median images were generated by aggregating co-registered grayscale ultrasound images acquired by each transducer and computing median intensity values across the image stack. A pretrained ResNet-18 model was fine-tuned on a dataset consisting of clinical median images from linear and curvilinear transducers. The dataset was divided into training, validation, and testing subsets, ensuring no overlap between training and test transducers. To assess generalizability, the model was also evaluated on an independent test set of 396 phantom-validated images from linear and curvilinear transducers. RESULTS: The model achieved 100% accuracy on the first dataset's test set. On the independent test set, it attained 87.4% accuracy with high sensitivity (0.84) and specificity (0.88), demonstrating robust generalization. Occlusion sensitivity maps confirmed the model's attention to uniformity artifact regions. CONCLUSION: The deep learning framework using clinical median images demonstrated robust performance across several linear and curvilinear transducer models. It could be integrated into the clinical QC workflow by automating artifact detection in an effective and timely manner. In our practice, it can flag median images classified as artifact-present, reducing human review time by approximately 80% while preserving detection accuracy.

Journal of Ultrasound in Medicine
Mayo Clinic (US), Johns Hopkins University (US)
Openalex Percentile: Top 1%
Ultrasound Imaging and Elastography
24.37
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.