Cryogenic etching device based on the conductive cooling method using liquid nitrogen bath
We developed a cryogenic etching device capable of precise temperature control utilizing liquid nitrogen conductive cooling. This device achieves electrode temperatures as low as -170 °C and enables real-time electrode temperature control, providing a reliable alternative to existing cryogenic etching techniques. To demonstrate the performance of our device, we conducted etching experiments on trench-patterned SiO2 thin films, successfully achieving anisotropic etching under cryogenic conditions. Compared with conventional methods such as flowing liquid nitrogen and chillers, our apparatus has a simple electrode structure, eliminates the risk of coolant leakage, and enables real-time temperature control. With these advantages, it is expected that our device based on the liquid nitrogen conductive cooling can be widely used in cryogenic etching research.
Authors
- Kil-Byoung Chai
- Eun-Beom Choi
- Woongsub Song
- Hyo-Chang Lee
Institutions
- Daejeon University (KR)
- Korea Atomic Energy Research Institute (KR)
- Korea Aerospace University (KR)
- Korea University of Science and Technology (KR)
Publication Details
- Journal
- Review of Scientific Instruments
- Published
- 2026-04-01
- DOI
- https://doi.org/10.1063/5.0305079
- Primary Topic
- Optical Coatings and Gratings
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Ministry of Trade, Industry and Energy
- Ministry of Science and ICT, South Korea