Verification of a Fluid–Structure Interaction Model for Aortic Stenosis Through Comparison With In Vitro Experiments

Aortic stenosis (AS) severity is typically assessed by measuring pressure drop across the aortic valve in rest. However, this flow-dependent metric is influenced by patients' cardiac function, complicating clinical decision-making on valve replacement. Moreover, assessment during rest does not reflect valvular dynamics under higher flow rates (e.g., during exercise), and may underestimate severity in moderate AS patients. Patient-specific Fluid-Structure Interaction (FSI) modelling offers a promising, non-invasive method to simulate valve dynamics under varying conditions, independent of pressure exerted by the left ventricle. Therefore, this study aimed to experimentally verify an aortic stenosis FSI model using a patient-specific aortic valve geometry, including calcifications, across flow conditions ranging from rest to exercise. To achieve this goal, in vitro experiments were conducted using a mock-loop circulatory system with patient-specific silicone rubber valve models, both calcified and non-calcified. These experiments were replicated in FSI simulations. Overall, good agreement was observed between simulated and experimental results for the non-calcified valve in terms of mean transvalvular flow (5% error on average) and aortic valve area (AVA) (10% error on average). Discrepancies were more pronounced in the calcified valve due to added complexity and uncertainty introduced by calcifications (8% and 7% error on average for mean transvalvular flow and AVA, respectively). For clinical implementation, two key challenges remain: (1) developing efficient and reliable methods to estimate valve leaflet material properties and pre-stress, and (2) verifying the model against a broader range of in vitro data, followed by validating it against real, clinical data. Despite the remaining challenges, this study demonstrated the feasibility of using FSI models as a complementary, non-invasive tool to assess AS severity and support clinical decision-making on valve replacement.

Authors

Institutions

Publication Details

Journal
International Journal for Numerical Methods in Biomedical Engineering
Published
2026-03-31
DOI
https://doi.org/10.1002/cnm.70169
Primary Topic
Cardiac Valve Diseases and Treatments
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Verification of a Fluid–Structure Interaction Model for Aortic Stenosis Through Comparison With In Vitro Experiments

Sabine Verstraeten, Nils Karajan, Roel Meiburg, W. Huberts et al.
International Journal for Numerical Methods in Biomedical Engineering
Cardiac Valve Diseases and Treatments
article

Verification of a Fluid–Structure Interaction Model for Aortic Stenosis Through Comparison With In Vitro Experiments

Sabine Verstraeten, Nils Karajan, Roel Meiburg, W. Huberts, M.C.M. Rutten, Marcel van’t Veer, P A L Tonino, F. N. van de Vosse, Koen Janssens, Martijn Hoeijmakers
article en

Abstract

Aortic stenosis (AS) severity is typically assessed by measuring pressure drop across the aortic valve in rest. However, this flow-dependent metric is influenced by patients' cardiac function, complicating clinical decision-making on valve replacement. Moreover, assessment during rest does not reflect valvular dynamics under higher flow rates (e.g., during exercise), and may underestimate severity in moderate AS patients. Patient-specific Fluid-Structure Interaction (FSI) modelling offers a promising, non-invasive method to simulate valve dynamics under varying conditions, independent of pressure exerted by the left ventricle. Therefore, this study aimed to experimentally verify an aortic stenosis FSI model using a patient-specific aortic valve geometry, including calcifications, across flow conditions ranging from rest to exercise. To achieve this goal, in vitro experiments were conducted using a mock-loop circulatory system with patient-specific silicone rubber valve models, both calcified and non-calcified. These experiments were replicated in FSI simulations. Overall, good agreement was observed between simulated and experimental results for the non-calcified valve in terms of mean transvalvular flow (5% error on average) and aortic valve area (AVA) (10% error on average). Discrepancies were more pronounced in the calcified valve due to added complexity and uncertainty introduced by calcifications (8% and 7% error on average for mean transvalvular flow and AVA, respectively). For clinical implementation, two key challenges remain: (1) developing efficient and reliable methods to estimate valve leaflet material properties and pre-stress, and (2) verifying the model against a broader range of in vitro data, followed by validating it against real, clinical data. Despite the remaining challenges, this study demonstrated the feasibility of using FSI models as a complementary, non-invasive tool to assess AS severity and support clinical decision-making on valve replacement.

International Journal for Numerical Methods in Biomedical EngineeringVol. 42(4)
Synopsys (Switzerland) (CH), Radboud University Nijmegen (NL), Catharina Ziekenhuis (NL), Synopsys (Germany) (DE), Eindhoven University of Technology (NL), University of Amsterdam (NL)
Horizon 2020 Framework Programme
Openalex Percentile: Top 29%
Cardiac Valve Diseases and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.