Assessing the diagnostic accuracy of synovial fluid send-out laboratory panels for suspected knee periprosthetic joint infection

Aims: Although synovial fluid send-out laboratory panels are increasingly used in the diagnosis of periprosthetic joint infection (PJI), the impact of transportation delays on the accuracy of synovial fluid white blood cell (WBC) count and polymorphonuclear neutrophil percentage (%PMN) remains unclear. The aim of this study was to assess the accuracy of two send-out synovial panels compared with institutional synovial WBC count and differential in patients evaluated for total knee arthroplasty (TKA) PJI. Methods: Between October 2015 and July 2023, all patients with a suspect TKA PJI with both an institutional synovial cell count and a corresponding next-generation sequencing (NGS) or α-defensin result, derived from separate aliquots of the same aspirate, were assessed. Synovial WBC count, %PMN, synovial CRP, α-defensin, serum CRP, serum ESR, and D-dimer were extracted from institutional and send-out laboratory results. The concordance of synovial WBC count and %PMN values was assessed using the 2018 International Consensus Meeting (ICM) thresholds for chronic PJI. Sensitivity, specificity, positive predictive value, and negative predictive value were calculated to evaluate the diagnostic performance of the institutional and send-out analyses. Results: In total, 212 TKAs (44 (20.8%) next-generation sequencing (NGS), 168 (79.2%) α-defensin) were identified. Relative to institutional analysis, send-out WBC counts were significantly lower (NGS -58.6% (SD 39.2%), α-defensin -4.4% (SD 30.1%); p < 0.001), whereas %PMN was similar for both (NGS -0.1%, α-defensin 1.2%; p = 0.365). Despite these differences, only four (1.9%) send-out WBC counts and nine (4.2%) send-out %PMN values altered ICM minor criteria thresholds, switching from 'Infected' to 'Possibly infected' only once (0.5%). The WBC and %PMN counts from the NGS panel had a sensitivity of 0.923 and a specificity of 1.000, while the α-defensin panel had a sensitivity of 0.970 and a specificity of 0.857 for diagnosing PJI. Conclusion: Synovial fluid cell count and differential values obtained from NGS and α-defensin panels are reliable for diagnosing chronic PJI.

Authors

Institutions

Publication Details

Journal
The Bone & Joint Journal
Published
2026-04-01
DOI
https://doi.org/10.1302/0301-620x.108b4.bjj-2024-1345.r2
Primary Topic
Orthopedic Infections and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessing the diagnostic accuracy of synovial fluid send-out laboratory panels for suspected knee periprosthetic joint infection

Sagar Telang, Cory K. Mayfield, Andrew Dobitsch, Arjun Aron et al.
The Bone & Joint Journal
Orthopedic Infections and Treatments
article

Assessing the diagnostic accuracy of synovial fluid send-out laboratory panels for suspected knee periprosthetic joint infection

Sagar Telang, Cory K. Mayfield, Andrew Dobitsch, Arjun Aron, Ryan Palmer, Nathanael D. Heckmann
article en

Abstract

Aims: Although synovial fluid send-out laboratory panels are increasingly used in the diagnosis of periprosthetic joint infection (PJI), the impact of transportation delays on the accuracy of synovial fluid white blood cell (WBC) count and polymorphonuclear neutrophil percentage (%PMN) remains unclear. The aim of this study was to assess the accuracy of two send-out synovial panels compared with institutional synovial WBC count and differential in patients evaluated for total knee arthroplasty (TKA) PJI. Methods: Between October 2015 and July 2023, all patients with a suspect TKA PJI with both an institutional synovial cell count and a corresponding next-generation sequencing (NGS) or α-defensin result, derived from separate aliquots of the same aspirate, were assessed. Synovial WBC count, %PMN, synovial CRP, α-defensin, serum CRP, serum ESR, and D-dimer were extracted from institutional and send-out laboratory results. The concordance of synovial WBC count and %PMN values was assessed using the 2018 International Consensus Meeting (ICM) thresholds for chronic PJI. Sensitivity, specificity, positive predictive value, and negative predictive value were calculated to evaluate the diagnostic performance of the institutional and send-out analyses. Results: In total, 212 TKAs (44 (20.8%) next-generation sequencing (NGS), 168 (79.2%) α-defensin) were identified. Relative to institutional analysis, send-out WBC counts were significantly lower (NGS -58.6% (SD 39.2%), α-defensin -4.4% (SD 30.1%); p < 0.001), whereas %PMN was similar for both (NGS -0.1%, α-defensin 1.2%; p = 0.365). Despite these differences, only four (1.9%) send-out WBC counts and nine (4.2%) send-out %PMN values altered ICM minor criteria thresholds, switching from 'Infected' to 'Possibly infected' only once (0.5%). The WBC and %PMN counts from the NGS panel had a sensitivity of 0.923 and a specificity of 1.000, while the α-defensin panel had a sensitivity of 0.970 and a specificity of 0.857 for diagnosing PJI. Conclusion: Synovial fluid cell count and differential values obtained from NGS and α-defensin panels are reliable for diagnosing chronic PJI.

The Bone & Joint JournalVol. 108-B(4)
University of Southern California (US), Yale University (US), University of California Davis Medical Center (US)
Good health and well-being
Openalex Percentile: Top 36%
Orthopedic Infections and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.