A Short-Lived Isotope Counting System (SLICS) to measure reaction cross sections producing light, beta-decaying nuclei on OMEGA
A new activation detector optimized for short-lived isotopes was designed, installed, and tested on OMEGA. The Short-Lived Isotope Counting System (SLICS) can be positioned as close as 35 cm from the OMEGA target chamber center in a ten-inch-manipulator. Stable target samples can be mounted on the front end of the detector assembly to undergo triton-, deuteron-, or neutron-induced reactions that produce short-lived reaction products that beta decay. The emitted betas are detected in a phoswich configuration to eliminate the gamma background. The detector is optimized for isotopes with half-lives from 150 ms to 20 s, filling a gap in OMEGA's current activation detectors for isotopes with half-lives >5 min. In a pilot study, SLICS achieved an efficiency of 24.5% ± 2.0% for counting 8Li, with a total of 505.8 ± 4.4 qualifying events. Using the existing triton laser-ion acceleration platform on OMEGA, this detector can be used to measure rarely studied tritium-induced reactions on light nuclei with an estimated uncertainty of 12%. The design of the detector and its performance during a first experiment employing deuterium as a surrogate will be discussed.
Authors
- George A. Marcus
- B. Stanley
- B. Wager
- S. J. Padalino
- A. K. Schwemmlein
- P. Gastis
- M. J. Rosenberg
- M. Yuly
- C. J. Forrest
Institutions
- Los Alamos National Laboratory (US)
- Houghton University (US)
- SUNY Geneseo (US)
- Applied Energetics (United States) (US)
Publication Details
- Journal
- Review of Scientific Instruments
- Published
- 2026-04-01
- DOI
- https://doi.org/10.1063/5.0316817
- Primary Topic
- Nuclear physics research studies
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- National Nuclear Security Administration
- Laboratory Directed Research and Development