Background: Radiation dose generated from computed tomography (CT) has drawn more attention to diagnostic radiology. It is a known fact that the risk of radiation-induced cancer is increasing, thereby necessitating the optimisation of dose in CT protocols. This study focused on determining the radiation dose of CT scans for the abdomen and lung using a 64-slice CT scanner to evaluate their correlation with body mass index (BMI). The objective of the study was to critically evaluate the relationship between BMI and radiation dose metrics in both CT lung and CT abdomen examinations. Methods: Data from 106 patients who underwent CT lung and CT abdomen examinations at an advanced diagnostic center were retrospectively analysed. The volume CT dose index (CTDIvol), dose-length product (DLP), the scan range, and skin to skin antero-posterior (AP) and lateral (LAT) of the patients were documented for further analysis. Effective dose (E) and size-specific dose estimate (SSDE) were also computed. Results: The mean BMI for CT lung was recorded as 24.85 (5.65). However, the correlation between BMI and the dose metrics (SSDE, E, DLP, and CTDIvol) was not significant, with correlation coefficients of 0.1278, 0.047, 0.047, and 0.1147, respectively. In contrast, the BMI for CT abdomen scans showed a moderate correlation with E (0.5898), SSDE (0.6288), DLP (0.5898), and CTDIvol (0.612). The results demonstrate that BMI can be used as a radiation dose metric in the case of CT abdomen scans, but has no influence on CT lung scans. Conclusion: These results further suggest that BMI could provide radiation dose analysis, which in turn leads to optimisation of CT scan parameters.

Authors

Publication Details

Journal
Universiti Putra Malaysia Institutional Repository (Universiti Putra Malaysia)
Published
2025-08-30
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Radiation dose of abdominal and lung computed tomography (CT) based on body mass index (BMI) as an indicator

Farouk Kabir Umar, Nor Azura Muhammad, Muhammad Kabeer Sulayman, Bashiru Lukuman et al.
Universiti Putra Malaysia Institutional Repository (Universiti Putra Malaysia)
Radiation Dose and Imaging
article

Radiation dose of abdominal and lung computed tomography (CT) based on body mass index (BMI) as an indicator

Farouk Kabir Umar, Nor Azura Muhammad, Muhammad Kabeer Sulayman, Bashiru Lukuman, Mazliana Ahmad Kamarudin, Juliana Mohd Radzi, Muhammad Khalis Karim, Che Azura Abdullah, Izdihar Kamal
article en

Abstract

Background: Radiation dose generated from computed tomography (CT) has drawn more attention to diagnostic radiology. It is a known fact that the risk of radiation-induced cancer is increasing, thereby necessitating the optimisation of dose in CT protocols. This study focused on determining the radiation dose of CT scans for the abdomen and lung using a 64-slice CT scanner to evaluate their correlation with body mass index (BMI). The objective of the study was to critically evaluate the relationship between BMI and radiation dose metrics in both CT lung and CT abdomen examinations. Methods: Data from 106 patients who underwent CT lung and CT abdomen examinations at an advanced diagnostic center were retrospectively analysed. The volume CT dose index (CTDIvol), dose-length product (DLP), the scan range, and skin to skin antero-posterior (AP) and lateral (LAT) of the patients were documented for further analysis. Effective dose (E) and size-specific dose estimate (SSDE) were also computed. Results: The mean BMI for CT lung was recorded as 24.85 (5.65). However, the correlation between BMI and the dose metrics (SSDE, E, DLP, and CTDIvol) was not significant, with correlation coefficients of 0.1278, 0.047, 0.047, and 0.1147, respectively. In contrast, the BMI for CT abdomen scans showed a moderate correlation with E (0.5898), SSDE (0.6288), DLP (0.5898), and CTDIvol (0.612). The results demonstrate that BMI can be used as a radiation dose metric in the case of CT abdomen scans, but has no influence on CT lung scans. Conclusion: These results further suggest that BMI could provide radiation dose analysis, which in turn leads to optimisation of CT scan parameters.

Universiti Putra Malaysia Institutional Repository (Universiti Putra Malaysia)
Good health and well-being
Openalex Percentile: Top 25%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Radiation dose of abdominal and lung computed tomography (CT) based on body mass index (BMI) as an indicator — Farouk Kabir Umar, Nor Azura Muhammad, et al. · Universiti Putra Malaysia Institutional Repository (Universiti Putra Malaysia) (2025) | TGRS Research Map | TGRS