Real‐time energy measurement of clinical carbon ion beams using a cross‐correlation time‐of‐flight method with parallel‐plate chambers

Abstract Background In carbon‐ion radiotherapy (CIRT), the beam energy determines both the particle range and the overall dosimetric quality. Range‐verification QA devices such as Zebra and Giraffe, which are based on multilayer ionization chambers (MLICs), can verify the range but only under dedicated QA conditions, leaving any energy deviations introduced by nozzle components undetected in real time. In particular, nozzle structures such as ridge filters can broaden or modulate the energy spectrum, causing the effective energy delivered to the patient to differ from the nominal accelerator setting. These limitations highlight the need for a real‐time method capable of verifying the beam energy under actual clinical operating conditions. Purpose We proposed a TOF‐based beam‐energy measurement concept that leverages a cross‐correlation analysis of full detector waveforms. Compact and radiation‐hard parallel‐plate chambers (PPCs) were developed and evaluated, in contrast to prior TOF systems based on semiconductor detectors. Methods PPCs (2.5 cm diameter active area, 0.4 mm gas gap) were operated in CO 2 . Two detectors were mounted coaxially with detector separations of 22.5 and 46.3 cm. Experiments were performed at Yonsei Heavy‐ion Therapy Center (HITC) using four nominal energies (102.6, 140.4, 250.3, 430 MeV/nucleon) and three intensities, covering the clinically interesting ranges. Signals were digitized with a 1 GHz bandwidth oscilloscope. For each spill, paired waveforms were cross‐correlated, and peak times were refined by parabolic interpolation to determine TOF. Precision and accuracy were evaluated across energies, intensities, and detector separations. Results The PPCs operated stably for all beam conditions. Under pencil‐beam delivery and normalized to 1 s acquisitions, the timing precision of the mean TOF (standard error) remained within 1 ps for both detector separations, scaling with (N: number of TOF samples per acquisition) and not representing the single‐particle TOF resolution. Residuals between measured and theoretical TOF remained within 80 ps across energies and distances. After relativistic conversion from TOF to kinetic energy and then to water‐equivalent range, all deviations were within a 1 mm range shift, meeting the recommended clinical criteria for range verification. Conclusions We demonstrated that compact CO 2 ‐filled PPCs, operated as a TOF pair, can measure carbon‐ion beam energy across the clinically relevant range of energies (≈100–430 MeV/u) and intensities used in routine treatment delivery. We achieved sub‐picosecond timing precision on the TOF mean (standard error) per 1 s acquisition and submillimeter water‐equivalent range accuracy using a robust cross‐correlation analysis method. These results open the way to the integration of PPC‐based TOF monitoring to tighten beam‐delivery tolerances and improve the reliability and safety of carbon‐ion radiotherapy.

Authors

Institutions

Publication Details

Journal
Medical Physics
Published
2026-03-29
DOI
https://doi.org/10.1002/mp.70391
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Real‐time energy measurement of clinical carbon ion beams using a cross‐correlation time‐of‐flight method with parallel‐plate chambers

P. Fonte, Yongdo Yun, Na Hye Kwon, Woong Sub Koom et al.
Medical Physics
Radiation Therapy and Dosimetry
article

Real‐time energy measurement of clinical carbon ion beams using a cross‐correlation time‐of‐flight method with parallel‐plate chambers

P. Fonte, Yongdo Yun, Na Hye Kwon, Woong Sub Koom, N. Carolino, Ho Jin Kim, C H Kim, Sung Woon Choi, Chae‐Seon Hong, Luís Lopes, Dong Wook Kim, Soorim Han, Min Cheol Han, Ho Lee, Jin Sung Kim, Do Won Kim
article en

Abstract

Abstract Background In carbon‐ion radiotherapy (CIRT), the beam energy determines both the particle range and the overall dosimetric quality. Range‐verification QA devices such as Zebra and Giraffe, which are based on multilayer ionization chambers (MLICs), can verify the range but only under dedicated QA conditions, leaving any energy deviations introduced by nozzle components undetected in real time. In particular, nozzle structures such as ridge filters can broaden or modulate the energy spectrum, causing the effective energy delivered to the patient to differ from the nominal accelerator setting. These limitations highlight the need for a real‐time method capable of verifying the beam energy under actual clinical operating conditions. Purpose We proposed a TOF‐based beam‐energy measurement concept that leverages a cross‐correlation analysis of full detector waveforms. Compact and radiation‐hard parallel‐plate chambers (PPCs) were developed and evaluated, in contrast to prior TOF systems based on semiconductor detectors. Methods PPCs (2.5 cm diameter active area, 0.4 mm gas gap) were operated in CO 2 . Two detectors were mounted coaxially with detector separations of 22.5 and 46.3 cm. Experiments were performed at Yonsei Heavy‐ion Therapy Center (HITC) using four nominal energies (102.6, 140.4, 250.3, 430 MeV/nucleon) and three intensities, covering the clinically interesting ranges. Signals were digitized with a 1 GHz bandwidth oscilloscope. For each spill, paired waveforms were cross‐correlated, and peak times were refined by parabolic interpolation to determine TOF. Precision and accuracy were evaluated across energies, intensities, and detector separations. Results The PPCs operated stably for all beam conditions. Under pencil‐beam delivery and normalized to 1 s acquisitions, the timing precision of the mean TOF (standard error) remained within 1 ps for both detector separations, scaling with (N: number of TOF samples per acquisition) and not representing the single‐particle TOF resolution. Residuals between measured and theoretical TOF remained within 80 ps across energies and distances. After relativistic conversion from TOF to kinetic energy and then to water‐equivalent range, all deviations were within a 1 mm range shift, meeting the recommended clinical criteria for range verification. Conclusions We demonstrated that compact CO 2 ‐filled PPCs, operated as a TOF pair, can measure carbon‐ion beam energy across the clinically relevant range of energies (≈100–430 MeV/u) and intensities used in routine treatment delivery. We achieved sub‐picosecond timing precision on the TOF mean (standard error) per 1 s acquisition and submillimeter water‐equivalent range accuracy using a robust cross‐correlation analysis method. These results open the way to the integration of PPC‐based TOF monitoring to tighten beam‐delivery tolerances and improve the reliability and safety of carbon‐ion radiotherapy.

Medical PhysicsVol. 53(4)
Yonsei University (KR), LIP - Laboratory of Instrumentation and Experimental Particle Physics (PT), Polytechnic Institute of Coimbra (PT), Syneos Health (South Korea) (KR), INTRASOFT International (Luxembourg) (LU)
Affordable and clean energy
Openalex Percentile: Top 3%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.