Validating an Autoclave Cycle for Sterilization of Select Agent-Infected Murine Carcasses

Introduction: Infectious disease research conducted in the Duke Regional Biocontainment Laboratory, a biosafety level-3 facility, often includes work with mice. Infected carcasses must be properly decontaminated at the completion of each study and prior to removal from the facility. Working with Select Agents requires validated methods for complete inactivation of the agent before carcasses can be removed from registered spaces. Methods: Biological indicators were implanted into the abdomen of unfrozen mouse carcasses, then autoclaved using different cycle parameters and packaging strategies to determine the best approach to ensure complete sterilization. Results: Both 20-min and 30-min autoclave cycles at 122°C, with three pre-vacuum pulses, failed to consistently sterilize unfrozen mouse carcasses. A 45-min autoclave cycle with the same parameters did result in the successful sterilization of carcasses when packaged at two or four carcasses in an 8” × 12” autoclave bag. Conclusions: This study supports the utilization of an autoclave cycle (122°C sterilization temperature, 45-min sterilization time, three pre-vacuum pulses) to sterilize unfrozen mouse carcasses, up to four per bag.

Authors

Institutions

Publication Details

Journal
Applied Biosafety
Published
2026-03-28
DOI
https://doi.org/10.1177/15356760261433036
Primary Topic
Bacillus and Francisella bacterial research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validating an Autoclave Cycle for Sterilization of Select Agent-Infected Murine Carcasses

Kristina J. Riebe, Rose Asrican, Scott Alderman, J. Patrick Condreay
Applied Biosafety
Bacillus and Francisella bacterial research
article

Validating an Autoclave Cycle for Sterilization of Select Agent-Infected Murine Carcasses

Kristina J. Riebe, Rose Asrican, Scott Alderman, J. Patrick Condreay
article en

Abstract

Introduction: Infectious disease research conducted in the Duke Regional Biocontainment Laboratory, a biosafety level-3 facility, often includes work with mice. Infected carcasses must be properly decontaminated at the completion of each study and prior to removal from the facility. Working with Select Agents requires validated methods for complete inactivation of the agent before carcasses can be removed from registered spaces. Methods: Biological indicators were implanted into the abdomen of unfrozen mouse carcasses, then autoclaved using different cycle parameters and packaging strategies to determine the best approach to ensure complete sterilization. Results: Both 20-min and 30-min autoclave cycles at 122°C, with three pre-vacuum pulses, failed to consistently sterilize unfrozen mouse carcasses. A 45-min autoclave cycle with the same parameters did result in the successful sterilization of carcasses when packaged at two or four carcasses in an 8” × 12” autoclave bag. Conclusions: This study supports the utilization of an autoclave cycle (122°C sterilization temperature, 45-min sterilization time, three pre-vacuum pulses) to sterilize unfrozen mouse carcasses, up to four per bag.

Applied Biosafety
Duke University (US)
Responsible consumption and production
Openalex Percentile: Top 4%
Bacillus and Francisella bacterial research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.