Concurrent validation of a novel intraoperative navigation platform for total knee arthroplasty

This cadaveric study aimed to assess the accuracy of the Next-AR navigation platform in terms of implant alignment, kinematics, and collateral ligament elongations in comparison to gold standard measurement tools. Computed tomography of 7 fresh-frozen legs were acquired for patient-specific guide design, identification of bony landmarks and collateral ligaments to acquire real-time feedback for bone resection, alignment and collateral ligament elongations using Next-AR system. The specimens underwent medially-stabilized TKA (Medacta) based on the mechanical alignment technique. The difference between target and achieved implant alignment parameters was assessed. The relative change in tibiofemoral kinematics, medial collateral ligament- (MCL), and lateral collateral ligament (LCL) elongations acquired with the Next-AR system were compared to reference values acquired with bone-pin mounted motion trackers (Vicon, UK) while performing quasistatic squatting in an ex vivo knee simulator. The mean differences between achieved and target hip-knee-ankle angle, femoral varus, femoral flexion, femoral rotation, tibial varus, and tibial slope measured -1.5°, 0.5°, 0.07°, -0.14°, 0.79° and -0.36°, respectively, with no statistically significant differences found (p ≥ 0.12). Knee flexion (p=0.32) and valgus orientation (p=0.63), obtained from Next-AR and Vicon systems were comparable. Collateral ligament elongations derived from Next-AR exhibited similar values for the anterior bundle of MCL (p=0.18) and LCL in the anterior and middle bundles (p>0.3) obtained from the Vicon system. Our findings showed that the Next-AR system achieved postoperative implant alignment close to the planned targets, with small systematic deviation (mean bias ≤ 1.5°), limited total error (RMSE ≤ 2.3°), and variability of paired differences ≤ 1.87°, while additionally providing quantitative real-time assessment of tibiofemoral kinematics and collateral ligament elongations.

Authors

Institutions

Publication Details

Journal
Archives of Orthopaedic and Trauma Surgery
Published
2026-03-28
DOI
https://doi.org/10.1007/s00402-026-06265-9
Primary Topic
Total Knee Arthroplasty Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Concurrent validation of a novel intraoperative navigation platform for total knee arthroplasty

Orçun Taylan, Maciej Biały, Arne Van de Vyver, L. Scheys et al.
Archives of Orthopaedic and Trauma Surgery
Total Knee Arthroplasty Outcomes
article

Concurrent validation of a novel intraoperative navigation platform for total knee arthroplasty

Orçun Taylan, Maciej Biały, Arne Van de Vyver, L. Scheys, T. Louwagie, Margot Demeulenaere, Geert Peersman
article en

Abstract

This cadaveric study aimed to assess the accuracy of the Next-AR navigation platform in terms of implant alignment, kinematics, and collateral ligament elongations in comparison to gold standard measurement tools. Computed tomography of 7 fresh-frozen legs were acquired for patient-specific guide design, identification of bony landmarks and collateral ligaments to acquire real-time feedback for bone resection, alignment and collateral ligament elongations using Next-AR system. The specimens underwent medially-stabilized TKA (Medacta) based on the mechanical alignment technique. The difference between target and achieved implant alignment parameters was assessed. The relative change in tibiofemoral kinematics, medial collateral ligament- (MCL), and lateral collateral ligament (LCL) elongations acquired with the Next-AR system were compared to reference values acquired with bone-pin mounted motion trackers (Vicon, UK) while performing quasistatic squatting in an ex vivo knee simulator. The mean differences between achieved and target hip-knee-ankle angle, femoral varus, femoral flexion, femoral rotation, tibial varus, and tibial slope measured -1.5°, 0.5°, 0.07°, -0.14°, 0.79° and -0.36°, respectively, with no statistically significant differences found (p ≥ 0.12). Knee flexion (p=0.32) and valgus orientation (p=0.63), obtained from Next-AR and Vicon systems were comparable. Collateral ligament elongations derived from Next-AR exhibited similar values for the anterior bundle of MCL (p=0.18) and LCL in the anterior and middle bundles (p>0.3) obtained from the Vicon system. Our findings showed that the Next-AR system achieved postoperative implant alignment close to the planned targets, with small systematic deviation (mean bias ≤ 1.5°), limited total error (RMSE ≤ 2.3°), and variability of paired differences ≤ 1.87°, while additionally providing quantitative real-time assessment of tibiofemoral kinematics and collateral ligament elongations.

Archives of Orthopaedic and Trauma SurgeryVol. 146(1)
Universitair Ziekenhuis Leuven (BE), Province of Antwerp (BE), KU Leuven (BE)
Life below water
Openalex Percentile: Top 3%
Total Knee Arthroplasty Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.