Agreement between a swept-source optical coherence tomography and dual Scheimpflug Placido analyser in healthy, keratoconus suspect and keratoconus eyes

To evaluate the agreement and interchangeability of anterior and posterior corneal parameters measured by the swept-source optical coherence tomography–based ANTERION and the dual Scheimpflug–Placido GALILEI G4 in normal, keratoconus suspect (KS), and keratoconus (KC) eyes. In this cross-sectional study, 221 eyes (129 normal, 40 KS, 52 KC) were examined using ANTERION and GALILEI G4 in a single session. Anterior and posterior keratometry (K1, K2, K Max), best-fit sphere (BFS), central corneal thickness (CCT), and minimum corneal thickness (MCT) were analysed. Inter-device agreement was assessed using paired statistical tests and Bland–Altman analysis. In normal eyes, most parameters showed statistically significant mean differences with wide limits of agreement (LoA). In KS eyes, no significant differences were observed in each parameter. In KC eyes, mean differences were minimal, though LoAs were wide, particularly for anterior keratometry and pachymetry; only MCT showed a statistically significant difference. Pachymetric measurements consistently differed between devices, with ANTERION generally reporting smaller values. The wide LoAs between ANTERION and GALILEI indicate that the devices are not interchangeable, particularly for longitudinal monitoring in keratoconus. Consistent use of a single device is recommended in clinical practice.

Authors

Institutions

Publication Details

Journal
BMC Ophthalmology
Published
2026-03-28
DOI
https://doi.org/10.1186/s12886-026-04730-y
Primary Topic
Corneal surgery and disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Agreement between a swept-source optical coherence tomography and dual Scheimpflug Placido analyser in healthy, keratoconus suspect and keratoconus eyes

Adzura Salam, Assoc. Prof. Dr. Khairidzan Mohd Kamal, Muhammad Adri Mohamed Shafit, Mohd-Asyraaf Abdul-Kadir
BMC Ophthalmology
Corneal surgery and disorders
article

Agreement between a swept-source optical coherence tomography and dual Scheimpflug Placido analyser in healthy, keratoconus suspect and keratoconus eyes

Adzura Salam, Assoc. Prof. Dr. Khairidzan Mohd Kamal, Muhammad Adri Mohamed Shafit, Mohd-Asyraaf Abdul-Kadir
article en

Abstract

To evaluate the agreement and interchangeability of anterior and posterior corneal parameters measured by the swept-source optical coherence tomography–based ANTERION and the dual Scheimpflug–Placido GALILEI G4 in normal, keratoconus suspect (KS), and keratoconus (KC) eyes. In this cross-sectional study, 221 eyes (129 normal, 40 KS, 52 KC) were examined using ANTERION and GALILEI G4 in a single session. Anterior and posterior keratometry (K1, K2, K Max), best-fit sphere (BFS), central corneal thickness (CCT), and minimum corneal thickness (MCT) were analysed. Inter-device agreement was assessed using paired statistical tests and Bland–Altman analysis. In normal eyes, most parameters showed statistically significant mean differences with wide limits of agreement (LoA). In KS eyes, no significant differences were observed in each parameter. In KC eyes, mean differences were minimal, though LoAs were wide, particularly for anterior keratometry and pachymetry; only MCT showed a statistically significant difference. Pachymetric measurements consistently differed between devices, with ANTERION generally reporting smaller values. The wide LoAs between ANTERION and GALILEI indicate that the devices are not interchangeable, particularly for longitudinal monitoring in keratoconus. Consistent use of a single device is recommended in clinical practice.

BMC Ophthalmology
Universiti Malaysia Sarawak (MY), International Islamic University Malaysia (MY), Sultan Zainal Abidin University (MY)
Openalex Percentile: Top 4%
Corneal surgery and disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Agreement between a swept-source optical coherence tomography and dual Scheimpflug Placido analyser in healthy, keratoconus suspect and keratoconus eyes — Adzura Salam, Assoc. Prof. Dr. Khairidzan Mohd Kamal, et al. · BMC Ophthalmology (2026) | TGRS Research Map | TGRS