Non-gas-phase AgF-mediated conversion combined with vapour sampling for field analysis of V-agents and selected pesticides using portable GC/MS

• Field-deployable AgF-mediated wipe-then-cleave analysis method for V-agents • Enables direct gas-phase detection via fluoridate formation from surface samples • Method provides strong analytical responses without liquid sample preparation • Dual operational approaches for trace- and high-load contamination • Validated with HAPSITE® portable GC/MS on multiple interior materials Rapid and reliable field identification of persistent chemical warfare agents remains a major challenge for CBRN responders. The low volatility and strong surface adsorption of V-type nerve agents severely limit the effectiveness of vapour-based detection techniques such as portable GC/MS systems. To address this, a wipe-then-cleave method based on AgF-mediated conversion was developed and validated for field analysis of V-agents and selected organophosphorus pesticides. The approach enables direct AgF cleavage of the collected analyte on the sampling swab, generating volatile fluoridate products suitable for gas-phase detection without prior extraction or sample preparation. The reaction proceeds rapidly under ambient conditions, allowing analysis within minutes. The method was evaluated using a HAPSITE® Smart PLUS portable GC/MS across representative interior materials. Two operational modes were established: a high-sensitivity mode for trace detection and a high-load mode for visible contamination where detector saturation must be avoided. Surface loadings of 3 µg VX, Russian VX, demeton-S-methyl, and isoprobenphos produced high responses, automatically identified with match factors ≥96/100. Limits of detection ranged from 20-200 ng on glass surfaces (MF≥85), depending on analyte. The high-load mode achieved 100 % identification with match factors ≥99 for 1-mg deposits. The workflow is simple, requires minimal equipment, and is operationally feasible for reliable on-site confirmation of V-type agents using portable GC/MS during CBRN field operations. Although certain structurally related organophosphorus compounds yield identical fluoridates, the method provides sufficient specificity to indicate a toxic precursor.

Authors

Institutions

Publication Details

Journal
Journal of Chromatography A
Published
2026-03-01
DOI
https://doi.org/10.1016/j.chroma.2026.466942
Primary Topic
Pesticide Exposure and Toxicity
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Non-gas-phase AgF-mediated conversion combined with vapour sampling for field analysis of V-agents and selected pesticides using portable GC/MS

Roger Magnusson, Kristina Arnoldsson, Anders Östin
Journal of Chromatography A
Pesticide Exposure and Toxicity
article

Non-gas-phase AgF-mediated conversion combined with vapour sampling for field analysis of V-agents and selected pesticides using portable GC/MS

Roger Magnusson, Kristina Arnoldsson, Anders Östin
article en

Abstract

• Field-deployable AgF-mediated wipe-then-cleave analysis method for V-agents • Enables direct gas-phase detection via fluoridate formation from surface samples • Method provides strong analytical responses without liquid sample preparation • Dual operational approaches for trace- and high-load contamination • Validated with HAPSITE® portable GC/MS on multiple interior materials Rapid and reliable field identification of persistent chemical warfare agents remains a major challenge for CBRN responders. The low volatility and strong surface adsorption of V-type nerve agents severely limit the effectiveness of vapour-based detection techniques such as portable GC/MS systems. To address this, a wipe-then-cleave method based on AgF-mediated conversion was developed and validated for field analysis of V-agents and selected organophosphorus pesticides. The approach enables direct AgF cleavage of the collected analyte on the sampling swab, generating volatile fluoridate products suitable for gas-phase detection without prior extraction or sample preparation. The reaction proceeds rapidly under ambient conditions, allowing analysis within minutes. The method was evaluated using a HAPSITE® Smart PLUS portable GC/MS across representative interior materials. Two operational modes were established: a high-sensitivity mode for trace detection and a high-load mode for visible contamination where detector saturation must be avoided. Surface loadings of 3 µg VX, Russian VX, demeton-S-methyl, and isoprobenphos produced high responses, automatically identified with match factors ≥96/100. Limits of detection ranged from 20-200 ng on glass surfaces (MF≥85), depending on analyte. The high-load mode achieved 100 % identification with match factors ≥99 for 1-mg deposits. The workflow is simple, requires minimal equipment, and is operationally feasible for reliable on-site confirmation of V-type agents using portable GC/MS during CBRN field operations. Although certain structurally related organophosphorus compounds yield identical fluoridates, the method provides sufficient specificity to indicate a toxic precursor.

Journal of Chromatography A
Swedish Defence Research Agency (SE)
Openalex Percentile: Top 4%
Pesticide Exposure and Toxicity
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.