Temperature and Viscosity Effects on Self-Expanding Nitinol Transcatheter Aortic Valve Performance

Abstract In-vitro testing is widely used to evaluate the hemodynamic performance of transcatheter aortic valves (TAVs). However, many studies assess self-expanding valves using blood analogs matched in viscosity at room temperature, despite the temperature-dependent mechanical behavior of nitinol stent frames. This study investigates the independent and combined effects of temperature and viscosity on the in-vitro performance assessment of a self-expanding TAV. Rheological characterization was performed to identify water-glycerin solutions with matched viscosities at room temperature and physiological temperature (37°C). A patient-specific aortic model was cast in silicone. A self-expanding Medtronic Evolut R valve was deployed within the compliant model and evaluated using a pulse-duplicating left heart simulator under physiological flow and pressure conditions. The transvalvular pressure gradient and effective orifice area (EOA) were calculated. Rheometry identified two fluid conditions with comparable viscosities at room and body temperatures. Heating the blood analog to 37°C resulted in significantly reduced transvalvular pressure gradient and increased EOA compared to the viscosity-matched room temperature condition (p<0.0001). At physiological body temperature, the higher viscosity fluid (42% glycerin) yielded further improvements in valve performance relative to the lower viscosity fluid(38% glycerin), with lower pressure drop and larger EOA (p<0.0001). These results demonstrate that temperature dependent effects influence the hemodynamic performance of self-expanding nitinol TAVs beyond viscosity matching alone. Combined consideration of temperature and viscosity is therefore necessary to ensure physiologically relevant in-vitro assessment of self-expanding TAVs.

Authors

Institutions

Publication Details

Journal
Journal of Medical Devices
Published
2026-03-27
DOI
https://doi.org/10.1115/1.4071539
Primary Topic
Cardiac Valve Diseases and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Temperature and Viscosity Effects on Self-Expanding Nitinol Transcatheter Aortic Valve Performance

Z. Zhang, Bruce P. Lee, Brennan Vogl, Hoda Hatoum et al.
Journal of Medical Devices
Cardiac Valve Diseases and Treatments
article

Temperature and Viscosity Effects on Self-Expanding Nitinol Transcatheter Aortic Valve Performance

Z. Zhang, Bruce P. Lee, Brennan Vogl, Hoda Hatoum, Ghasaq Saleh, Ahmad Bshennaty, Mohamad Adnan Alkhouli
article en

Abstract

Abstract In-vitro testing is widely used to evaluate the hemodynamic performance of transcatheter aortic valves (TAVs). However, many studies assess self-expanding valves using blood analogs matched in viscosity at room temperature, despite the temperature-dependent mechanical behavior of nitinol stent frames. This study investigates the independent and combined effects of temperature and viscosity on the in-vitro performance assessment of a self-expanding TAV. Rheological characterization was performed to identify water-glycerin solutions with matched viscosities at room temperature and physiological temperature (37°C). A patient-specific aortic model was cast in silicone. A self-expanding Medtronic Evolut R valve was deployed within the compliant model and evaluated using a pulse-duplicating left heart simulator under physiological flow and pressure conditions. The transvalvular pressure gradient and effective orifice area (EOA) were calculated. Rheometry identified two fluid conditions with comparable viscosities at room and body temperatures. Heating the blood analog to 37°C resulted in significantly reduced transvalvular pressure gradient and increased EOA compared to the viscosity-matched room temperature condition (p<0.0001). At physiological body temperature, the higher viscosity fluid (42% glycerin) yielded further improvements in valve performance relative to the lower viscosity fluid(38% glycerin), with lower pressure drop and larger EOA (p<0.0001). These results demonstrate that temperature dependent effects influence the hemodynamic performance of self-expanding nitinol TAVs beyond viscosity matching alone. Combined consideration of temperature and viscosity is therefore necessary to ensure physiologically relevant in-vitro assessment of self-expanding TAVs.

Journal of Medical Devices
Michigan Technological University (US), Mayo Clinic in Arizona (US), Biocom (US)
Clean water and sanitation
Openalex Percentile: Top 5%
Cardiac Valve Diseases and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.