Modeling and estimation of ablation catheter contact force under blood flow disturbance
Real-time, accurate estimation of catheter tip-atrial tissue contact force (CF) is essential for safe, effective robot-assisted radiofrequency catheter ablation (RFCA). Conventional models often neglect intracardiac hemodynamics or reduce them to steady flow, missing cardiac-cycle velocity variations and pulsatile disturbances, thus causing large errors. This study proposes a CF estimation framework using cubic Bezier splines for catheter morphology, a sinusoidal velocity model for pulsatile blood flow, and a quasi-static Euler-Bernoulli beam moment equilibrium with distributed hydrodynamic loads. CF is inversely estimated via least-squares and the Moore-Penrose pseudoinverse. FE and in-vitro results show improved dynamic-condition accuracy and physiologically relevant intraoperative feedback.
Authors
- Lei Sun (ORCID: https://orcid.org/0000-0001-6662-1844)
- Zhe Hu (ORCID: https://orcid.org/0000-0001-7652-6939)
- Hanping Zhu
- Yifan Liang
Institutions
- Shanghai University of Engineering Science (CN)
Publication Details
- Journal
- Computer Methods in Biomechanics & Biomedical Engineering
- Published
- 2026-03-26
- DOI
- https://doi.org/10.1080/10255842.2026.2642808
- Primary Topic
- Soft Robotics and Applications
- Type
- article
- Field-Weighted Citation Impact