Performance of Wearable Pulse Oximetry During Controlled Hypoxia Induction: Instrument Validation Study
Both devices consistently overestimated SpO2, with accuracy declining notably during hypoxemia. The Apple Watch Series 7 mean bias suggests a likelihood of missing instances of hypoxemia, particularly at arterial oxygen saturation values below but close to 88%. Both the Apple Watch Series 7 and Masimo MightySat Rx exhibited Arms values exceeding the US Food and Drug Administration threshold under conditions of hypoxemia. While past studies have implicated high ODRs in increasing measurement error, we found no statistically significant relationship between ODR and measurement error for either device. Overall, our findings of SpO2 overestimation and high Arms values underscore the need for caution when interpreting oxygen saturation values from these devices. The small sample size and limited diversity in skin tone and age restrict the generalizability of our findings. Future studies should include larger and more diverse populations to evaluate the performance of wearable-based pulse oximetry.
Authors
- Jessilyn Dunn (ORCID: https://orcid.org/0000-0002-3241-8183)
- Yihang Jiang
- Satasuk Joy Bhosai
- David B. MacLeod (ORCID: https://orcid.org/0000-0002-5989-3961)
- Laurie D. Snyder (ORCID: https://orcid.org/0000-0002-5962-2184)
- Ali R. Roghanizad (ORCID: https://orcid.org/0000-0002-3461-5092)
- Connor Spies (ORCID: https://orcid.org/0000-0003-2375-4561)
- Will Ke Wang
- Ashley Burke
Institutions
- Duke University (US)
- Clinical Research Institute (US)
Publication Details
- Journal
- JMIR Formative Research
- Published
- 2026-03-27
- DOI
- https://doi.org/10.2196/85253
- Primary Topic
- Non-Invasive Vital Sign Monitoring
- Type
- article
- Field-Weighted Citation Impact
- 0.00