Accuracy, Repeatability, and Test–Retest Reliability of a Pressure Algometer for Pain Threshold and Tolerance in Sports, Exercise, and Rehabilitation Settings

Background: Pressure algometry is commonly used in sports, exercise, and rehabilitation settings to assess pain sensitivity and monitor neuromuscular status. Reliable and accessible devices are required for consistent assessment. This study evaluated the accuracy, agreement, repeatability, and test–retest reliability of the AMF-500 digital pressure algometer. Methods: Three independent studies were conducted. Study A assessed the agreement between the AMF-500 and a three-axis AMTI force plate during 30 controlled pressure trials. Study B compared pressure pain thresholds (lumbar paravertebral muscles, tibialis anterior, and thenar eminence) and lumbar pressure pain tolerance between the AMF-500 and the MED.DOR algometer in 27 healthy adults. Study C (n = 27 healthy adults) evaluated test–retest reliability across two sessions separated by 48 h. Agreement was assessed using Bland–Altman analyses, and intraclass correlation coefficients (ICC) was also applied. Standard error of measurement (SEM) and minimal detectable change (MDC) were also calculated. All pressure values were expressed in N/cm2. Results: In Study A, the AMF-500 slightly overestimated pressure compared with the force plate (19.34 ± 2.44 vs. 18.71 ± 2.49 N/cm2), with a mean bias of 0.63 N/cm2 and limits of agreement from 0.21 to 1.05 N/cm2, corresponding to a mean difference of approximately 3.4%. Despite this small systematic bias, agreement between devices was excellent (ICC = 0.99; 95% CI: 0.96–0.99), and no proportional bias was detected, indicating a small and consistent overestimation of pressure by the AMF-500. In Study B, no significant differences were observed between AMF-500 and MED.DOR for lumbar threshold, lumbar tolerance, tibialis anterior threshold, or thenar eminence threshold (all p > 0.05). Agreement between devices was good to excellent (ICC = 0.82–0.91). Bland–Altman analyses showed small mean biases (1.05–7.51 N/cm2), with proportional bias detected only for lumbar tolerance. In Study C, test–retest reliability for the AMF-500 ranged from moderate to good across sites (ICC = 0.69–0.88), comparable to MED.DOR (ICC = 0.63–0.88). SEM values for the AMF-500 ranged from 9.31 to 25.15 N/cm2, with higher variability observed for lumbar tolerance. Conclusions: The AMF-500 demonstrated acceptable accuracy, agreement, and reliability when compared with a laboratory force plate and an established clinical algometer. These findings can support its use as a low-cost tool for pressure pain assessment in sports, exercise, and rehabilitation contexts.

Authors

Institutions

Publication Details

Journal
Applied Sciences
Published
2026-03-26
DOI
https://doi.org/10.3390/app16073210
Primary Topic
Musculoskeletal pain and rehabilitation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accuracy, Repeatability, and Test–Retest Reliability of a Pressure Algometer for Pain Threshold and Tolerance in Sports, Exercise, and Rehabilitation Settings

Jader Pereira de Farias Neto, Rodrigo Zacca, Leonardo Yung dos Santos Maciel, V. G. Oliveira Almeida et al.
Applied Sciences
Musculoskeletal pain and rehabilitation
article

Accuracy, Repeatability, and Test–Retest Reliability of a Pressure Algometer for Pain Threshold and Tolerance in Sports, Exercise, and Rehabilitation Settings

Jader Pereira de Farias Neto, Rodrigo Zacca, Leonardo Yung dos Santos Maciel, V. G. Oliveira Almeida, Valter Joviniano de Santana Filho
article en

Abstract

Background: Pressure algometry is commonly used in sports, exercise, and rehabilitation settings to assess pain sensitivity and monitor neuromuscular status. Reliable and accessible devices are required for consistent assessment. This study evaluated the accuracy, agreement, repeatability, and test–retest reliability of the AMF-500 digital pressure algometer. Methods: Three independent studies were conducted. Study A assessed the agreement between the AMF-500 and a three-axis AMTI force plate during 30 controlled pressure trials. Study B compared pressure pain thresholds (lumbar paravertebral muscles, tibialis anterior, and thenar eminence) and lumbar pressure pain tolerance between the AMF-500 and the MED.DOR algometer in 27 healthy adults. Study C (n = 27 healthy adults) evaluated test–retest reliability across two sessions separated by 48 h. Agreement was assessed using Bland–Altman analyses, and intraclass correlation coefficients (ICC) was also applied. Standard error of measurement (SEM) and minimal detectable change (MDC) were also calculated. All pressure values were expressed in N/cm2. Results: In Study A, the AMF-500 slightly overestimated pressure compared with the force plate (19.34 ± 2.44 vs. 18.71 ± 2.49 N/cm2), with a mean bias of 0.63 N/cm2 and limits of agreement from 0.21 to 1.05 N/cm2, corresponding to a mean difference of approximately 3.4%. Despite this small systematic bias, agreement between devices was excellent (ICC = 0.99; 95% CI: 0.96–0.99), and no proportional bias was detected, indicating a small and consistent overestimation of pressure by the AMF-500. In Study B, no significant differences were observed between AMF-500 and MED.DOR for lumbar threshold, lumbar tolerance, tibialis anterior threshold, or thenar eminence threshold (all p > 0.05). Agreement between devices was good to excellent (ICC = 0.82–0.91). Bland–Altman analyses showed small mean biases (1.05–7.51 N/cm2), with proportional bias detected only for lumbar tolerance. In Study C, test–retest reliability for the AMF-500 ranged from moderate to good across sites (ICC = 0.69–0.88), comparable to MED.DOR (ICC = 0.63–0.88). SEM values for the AMF-500 ranged from 9.31 to 25.15 N/cm2, with higher variability observed for lumbar tolerance. Conclusions: The AMF-500 demonstrated acceptable accuracy, agreement, and reliability when compared with a laboratory force plate and an established clinical algometer. These findings can support its use as a low-cost tool for pressure pain assessment in sports, exercise, and rehabilitation contexts.

Applied SciencesVol. 16(7)
Universidade do Porto (PT), Universidade Federal de Sergipe (BR), Universidad Adventista de Chile (CL)
Good health and well-being
Openalex Percentile: Top 6%
Musculoskeletal pain and rehabilitation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.