Whole-Body Dose of Patients Treated With ZAP-X Intracranial Gyroscopic Stereotactic Radiosurgery

During the delivery of radiosurgery treatments, a small dose is given unintentionally to the areas of the patient’s body outside the primary treatment volume. It is a common practice for radiotherapy platform developers to minimize the dose outside the primary radiation field by providing X-ray target shielding in the beam-defining collimator head of such systems. The amount of such doses was measured for the ZAP-X® delivery platform by designing a phantom-based set-up mimicking the patient and measuring the dose with an ionization chamber and OSLDs. Furthermore, during several clinical ZAP-X radiosurgery treatment deliveries, the dose received by patients was measured with film-based dosimeters. The results were determined as absolute values and as a percentage of the primary dose.

Authors

Publication Details

Journal
Cureus
Published
2026-03-26
DOI
https://doi.org/10.7759/cureus.c419
Primary Topic
Vascular Malformations Diagnosis and Treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Whole-Body Dose of Patients Treated With ZAP-X Intracranial Gyroscopic Stereotactic Radiosurgery

Georg A. Weidlich
Cureus
Vascular Malformations Diagnosis and Treatment
article

Whole-Body Dose of Patients Treated With ZAP-X Intracranial Gyroscopic Stereotactic Radiosurgery

Georg A. Weidlich
article en

Abstract

During the delivery of radiosurgery treatments, a small dose is given unintentionally to the areas of the patient’s body outside the primary treatment volume. It is a common practice for radiotherapy platform developers to minimize the dose outside the primary radiation field by providing X-ray target shielding in the beam-defining collimator head of such systems. The amount of such doses was measured for the ZAP-X® delivery platform by designing a phantom-based set-up mimicking the patient and measuring the dose with an ionization chamber and OSLDs. Furthermore, during several clinical ZAP-X radiosurgery treatment deliveries, the dose received by patients was measured with film-based dosimeters. The results were determined as absolute values and as a percentage of the primary dose.

Cureus
Good health and well-being
Openalex Percentile: Top 13%
Vascular Malformations Diagnosis and Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.