Mechanical Properties of Percutaneous Transluminal Coronary Angioplasty Balloon Catheters: A Bench Study.

Purpose This paper aims to categorize the mechanical properties of eight percutaneous transluminal coronary angioplasty (PTCA) balloon catheters and translate them into clinical needs.Method Objective bench tests quantified PTCA balloon catheters' mechanical properties, including tensile strength, kink resistance, bending, torsional behavior, friction, radio-opacity, pushability, and trackability. The results were compared against each other and supplemented by a survey of interventional cardiologists.Results Clinical needs with respect to deliverability, dilatation efficiency, and crossability were assessed for each catheter. Results indicate that SC Maverick2 excels in deliverability and crossability but lags in dilatation efficiency compared to NC catheters, though it is rated best to fulfill clinical needs. SC EasyT outperforms Maverick2 in dilatation efficiency but has inferior deliverability, similar to NC catheters. The NC balloons were comparable. However, Accuforce exhibits highest deliverability, Pantera LEO shows superior dilatation efficiency, NC Emerge shows best crossability, and Sapphire NC24 exhibits lowest performance. According to the survey, Accuforce, Sapphire NC24, and NC Trek are favored over NC Emerge and Pantera LEO. OPN NC offers limited deliverability but can treat lesions where standard NC catheters fail due to its unique rated burst pressure. Trackability and pushability estimates align better with survey results than those obtained from simulated use-case tests, except for Sapphire NC24 and OPN NC. Torquability measurements show discrepancies with survey ratings, indicating additional influences and rating challenges.Conclusion Our study comprehensively analyzes PTCA balloon catheters, emphasizing the importance of integrating mechanical and design attributes throughout development. Performance properties like trackability and pushability were measured, providing unbiased insights for comparison independent of individual practitioner preferences.

Authors

Institutions

Publication Details

Journal
Open Access CRIS of the University of Bern
Published
2026-03-22
DOI
https://doi.org/10.48620/96388
Primary Topic
Coronary Interventions and Diagnostics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Mechanical Properties of Percutaneous Transluminal Coronary Angioplasty Balloon Catheters: A Bench Study.

A. Haeberlin, M. Hersberger, R. Vogel, A. Zurbuchen et al.
Open Access CRIS of the University of Bern
Coronary Interventions and Diagnostics
article

Mechanical Properties of Percutaneous Transluminal Coronary Angioplasty Balloon Catheters: A Bench Study.

A. Haeberlin, M. Hersberger, R. Vogel, A. Zurbuchen, J. Burger, C. Amstutz, N. Fontaine, J Behr
article en

Abstract

Purpose This paper aims to categorize the mechanical properties of eight percutaneous transluminal coronary angioplasty (PTCA) balloon catheters and translate them into clinical needs.Method Objective bench tests quantified PTCA balloon catheters' mechanical properties, including tensile strength, kink resistance, bending, torsional behavior, friction, radio-opacity, pushability, and trackability. The results were compared against each other and supplemented by a survey of interventional cardiologists.Results Clinical needs with respect to deliverability, dilatation efficiency, and crossability were assessed for each catheter. Results indicate that SC Maverick2 excels in deliverability and crossability but lags in dilatation efficiency compared to NC catheters, though it is rated best to fulfill clinical needs. SC EasyT outperforms Maverick2 in dilatation efficiency but has inferior deliverability, similar to NC catheters. The NC balloons were comparable. However, Accuforce exhibits highest deliverability, Pantera LEO shows superior dilatation efficiency, NC Emerge shows best crossability, and Sapphire NC24 exhibits lowest performance. According to the survey, Accuforce, Sapphire NC24, and NC Trek are favored over NC Emerge and Pantera LEO. OPN NC offers limited deliverability but can treat lesions where standard NC catheters fail due to its unique rated burst pressure. Trackability and pushability estimates align better with survey results than those obtained from simulated use-case tests, except for Sapphire NC24 and OPN NC. Torquability measurements show discrepancies with survey ratings, indicating additional influences and rating challenges.Conclusion Our study comprehensively analyzes PTCA balloon catheters, emphasizing the importance of integrating mechanical and design attributes throughout development. Performance properties like trackability and pushability were measured, providing unbiased insights for comparison independent of individual practitioner preferences.

Open Access CRIS of the University of Bern
Precision for Medicine (United States) (US), Precision Research (United States) (US)
Openalex Percentile: Top 7%
Coronary Interventions and Diagnostics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.