Two-parameter analytical model for paralyzable radiation detectors

Certain radiation detectors become "paralyzed" at high input count rates. When applied to count rates close to the event discriminator working rate, the classic one-parameter dead time model fails. Here we present a corrected paralyzable detector model accounting for the event discriminator’s finite response time. This two-parameter analytical model, when compared to the experimental data from a commercial x-ray detector, gives an improved description of the input and output count rate relations. Furthermore, it can independently determine the discriminator response time and the pulse shaper dead time, critical parameters for understanding a detector’s performance. Finally, this model also provides a post-acquisition pile-up correction that greatly reduces artifacts in high-throughput spectra. In some situations, applying this model to optimize the acquisition and post-acquisition correction allows a user to acquire data an order of magnitude faster without compromising accuracy.

Authors

Publication Details

Journal
Physical Review Applied
Published
2026-11-18
DOI
https://doi.org/10.1103/plf3-jtkp
Citations
1
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
2.66

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Two-parameter analytical model for paralyzable radiation detectors

Yueyun Chen, Matthew H. Mecklenburg
1 citations
Physical Review Applied
Radiation Detection and Scintillator Technologies
2.66
article

Two-parameter analytical model for paralyzable radiation detectors

Yueyun Chen, Matthew H. Mecklenburg
article en
1 citations

Abstract

Certain radiation detectors become "paralyzed" at high input count rates. When applied to count rates close to the event discriminator working rate, the classic one-parameter dead time model fails. Here we present a corrected paralyzable detector model accounting for the event discriminator’s finite response time. This two-parameter analytical model, when compared to the experimental data from a commercial x-ray detector, gives an improved description of the input and output count rate relations. Furthermore, it can independently determine the discriminator response time and the pulse shaper dead time, critical parameters for understanding a detector’s performance. Finally, this model also provides a post-acquisition pile-up correction that greatly reduces artifacts in high-throughput spectra. In some situations, applying this model to optimize the acquisition and post-acquisition correction allows a user to acquire data an order of magnitude faster without compromising accuracy.

Physical Review Applied
National Science Foundation, University of California, Los Angeles, Division of Materials Research, California NanoSystems Institute
Openalex Percentile: Top 18%
Radiation Detection and Scintillator Technologies
2.66
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon