Validation of the HemoCue® HbA1c 501 System for Point-of-care HbA1c Estimation Using Venous and Capillary Blood: A Comparative Study with HPLC in Eastern India

Abstract Introduction: A valid point-of-care testing (POCT) method for estimating HbA1c values is required in field settings. We conducted this study to compare the validity of the HemoCue HbA1c 501 System (a POCT device) in estimating HbA1c among patients attending a tertiary healthcare facility, considering high-performance liquid chromatography (HPLC) for HbA1c estimation as the gold standard. Methods: This prospective observational study was conducted among the patients attending the outpatient department in a tertiary care hospital in eastern India. The HbA1c was estimated by HPLC and HemoCue ® HbA1c 501 System POCT device using both venous and capillary blood. The HbA1c values were categorized using a cutoff of ≥ 7% to calculate sensitivity, specificity, and positive and negative predictive values. Bland–Altman plot, Lin’s concordance correlation coefficient (CCC), intra-class correlation coefficient, and Kappa statistics were used to assess the validity of the POCT. Results: Lin’s CCC was 0.95 (95% CI: 0.93 to 0.96, P value < 0.001) for venous blood and 0.96 (CI: 0.95–0.97) for capillary blood using POCT. The sensitivity was 95.24%, the specificity was 92.70%, the positive predictive value was 85.71%, and the negative predictive value was 97.69% for HemoCue ® HbA1c 501 using venous blood for the detection of hyperglycemia. The sensitivity was 96.72%, the specificity was 97.12%, the positive predictive value was 93.65%, and the negative predictive value was 98.54% for HemoCue ® HbA1c 501 using capillary blood for the detection of hyperglycemia. Conclusion: HemoCue ® HbA1c has good sensitivity, specificity, almost perfect agreement, and substantial concordance with the gold-standard HPLC for HbA1c estimation.

Authors

Institutions

Publication Details

Journal
Indian Journal of Community Medicine
Published
2026-03-26
DOI
https://doi.org/10.4103/ijcm.ijcm_432_25
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validation of the HemoCue® HbA1c 501 System for Point-of-care HbA1c Estimation Using Venous and Capillary Blood: A Comparative Study with HPLC in Eastern India

Sarika Palepu, D. Pal, Sibasish Sahoo, A.C. Roy et al.
Indian Journal of Community Medicine
Clinical Laboratory Practices and Quality Control
article

Validation of the HemoCue® HbA1c 501 System for Point-of-care HbA1c Estimation Using Venous and Capillary Blood: A Comparative Study with HPLC in Eastern India

Sarika Palepu, D. Pal, Sibasish Sahoo, A.C. Roy, Kalyan Goswami, Farhad Ahamed, Jeevanmuktha Somashekara
article en

Abstract

Abstract Introduction: A valid point-of-care testing (POCT) method for estimating HbA1c values is required in field settings. We conducted this study to compare the validity of the HemoCue HbA1c 501 System (a POCT device) in estimating HbA1c among patients attending a tertiary healthcare facility, considering high-performance liquid chromatography (HPLC) for HbA1c estimation as the gold standard. Methods: This prospective observational study was conducted among the patients attending the outpatient department in a tertiary care hospital in eastern India. The HbA1c was estimated by HPLC and HemoCue ® HbA1c 501 System POCT device using both venous and capillary blood. The HbA1c values were categorized using a cutoff of ≥ 7% to calculate sensitivity, specificity, and positive and negative predictive values. Bland–Altman plot, Lin’s concordance correlation coefficient (CCC), intra-class correlation coefficient, and Kappa statistics were used to assess the validity of the POCT. Results: Lin’s CCC was 0.95 (95% CI: 0.93 to 0.96, P value < 0.001) for venous blood and 0.96 (CI: 0.95–0.97) for capillary blood using POCT. The sensitivity was 95.24%, the specificity was 92.70%, the positive predictive value was 85.71%, and the negative predictive value was 97.69% for HemoCue ® HbA1c 501 using venous blood for the detection of hyperglycemia. The sensitivity was 96.72%, the specificity was 97.12%, the positive predictive value was 93.65%, and the negative predictive value was 98.54% for HemoCue ® HbA1c 501 using capillary blood for the detection of hyperglycemia. Conclusion: HemoCue ® HbA1c has good sensitivity, specificity, almost perfect agreement, and substantial concordance with the gold-standard HPLC for HbA1c estimation.

Indian Journal of Community Medicine
All India Institute of Medical Sciences (IN), University of Kalyani (IN)
Openalex Percentile: Top 7%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.