Disintegration of the Second‐Generation Precice Bone Transport Nail During Removal Is Still an Issue: A Case Report

Background Intramedullary bone transport nails (BTNs) have become an increasingly common method for managing bone defects. The Precice BTN (Globus Medical, Inc.) was previously associated with mechanical failures including intramedullary disintegration during elective nail removal. These issues were reportedly resolved in the second generation BTN. Results Adhering to the CARE checklist, we describe a case involving a 23‐year‐old male with a posttraumatic bone defect following a motorcycle accident. He sustained a Gustilo–Anderson Type IIIB comminuted right fracture with a 5 cm bone defect. The leg was initially stabilized with external fixation. After several debridements, the fracture was internally stabilized using a regular trauma nail while the soft‐tissue defect was covered with a free flap and split‐thickness skin grafting. Subsequently, the 5 cm bone defect was managed using a second‐generation BTN. After 8 months, the patient underwent a planned nail removal according to the manufacturer′s instructions. Just as with the first‐generation BTN, the nail disintegrated within the medullary canal during the procedure and separated into multiple components. These internal components were cumbersomely removed using long forceps. Conclusion Mechanical failure, specifically intramedullary disintegration, still occurs with the second‐generation BTN upon nail removal. The retention plug of the first‐generation BTN should thus be used despite instructions from the manufacturer that the problem has been solved in the current generation BTN. Appropriate instruments to manage potential nail disintegration should be present. The case underlines the need for further design improvements in future generations of BTN.

Authors

Institutions

Publication Details

Journal
Case Reports in Orthopedics
Published
2026-01-01
DOI
https://doi.org/10.1155/cro/5629022
Primary Topic
Bone fractures and treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Disintegration of the Second‐Generation Precice Bone Transport Nail During Removal Is Still an Issue: A Case Report

Juozas Petruskevicius, Daniel Wæver, Jan Duedal Rölfing, Rikke Gravesen
Case Reports in Orthopedics
Bone fractures and treatments
article

Disintegration of the Second‐Generation Precice Bone Transport Nail During Removal Is Still an Issue: A Case Report

Juozas Petruskevicius, Daniel Wæver, Jan Duedal Rölfing, Rikke Gravesen
article en

Abstract

Background Intramedullary bone transport nails (BTNs) have become an increasingly common method for managing bone defects. The Precice BTN (Globus Medical, Inc.) was previously associated with mechanical failures including intramedullary disintegration during elective nail removal. These issues were reportedly resolved in the second generation BTN. Results Adhering to the CARE checklist, we describe a case involving a 23‐year‐old male with a posttraumatic bone defect following a motorcycle accident. He sustained a Gustilo–Anderson Type IIIB comminuted right fracture with a 5 cm bone defect. The leg was initially stabilized with external fixation. After several debridements, the fracture was internally stabilized using a regular trauma nail while the soft‐tissue defect was covered with a free flap and split‐thickness skin grafting. Subsequently, the 5 cm bone defect was managed using a second‐generation BTN. After 8 months, the patient underwent a planned nail removal according to the manufacturer′s instructions. Just as with the first‐generation BTN, the nail disintegrated within the medullary canal during the procedure and separated into multiple components. These internal components were cumbersomely removed using long forceps. Conclusion Mechanical failure, specifically intramedullary disintegration, still occurs with the second‐generation BTN upon nail removal. The retention plug of the first‐generation BTN should thus be used despite instructions from the manufacturer that the problem has been solved in the current generation BTN. Appropriate instruments to manage potential nail disintegration should be present. The case underlines the need for further design improvements in future generations of BTN.

Case Reports in OrthopedicsVol. 2026(1)
Aarhus University Hospital (DK), Regional Hospital Randers (DK)
Good health and well-being
Openalex Percentile: Top 10%
Bone fractures and treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Disintegration of the Second‐Generation Precice Bone Transport Nail During Removal Is Still an Issue: A Case Report — Juozas Petruskevicius, Daniel Wæver, et al. · Case Reports in Orthopedics (2026) | TGRS Research Map | TGRS