Oxygenator oxygen transfer performance during cardiopulmonary bypass of the Sorin <sup>®</sup> Inspire 6F <sup>©</sup>

BackgroundOxygenator oxygen transfer performance is typically measured by manufacturers under standardized, static laboratory conditions. However, these parameters do not adequately reflect the dynamic environment of clinical cardiopulmonary bypass (CPB). To align with modern goal-directed perfusion standards - specifically the maintenance of physiological PaO2 levels to reduce oxidative stress - perfusionists require a reliable tool to titrate FiO2 based on real-time oxygen demand. This study aimed to develop a clinical oxygen transfer function for the Sorin® Inspire 6F© oxygenator.MethodsWe retrospectively analyzed 420 blood gas samples and hemodynamic parameters from 54 adult patients undergoing CPB with the Sorin® Inspire 6F©. Oxygen transfer (VO2) was calculated via the Fick principle. Using the methodology established by Hamilton et al., a validation subset (<i>n</i> = 57) of samples within a target PaO2 range of 150 ± 5 mmHg was used to derive a specific correlation function between FiO2 and VO2 via linear regression analysis.ResultsA significant correlation was observed between FiO2 and oxygen transfer. For the validation subset, the derived oxygen transfer function for the Sorin® Inspire 6F© was VO<sub>2</sub> = 284.34 x FiO2 (R<sup>2</sup> = 0.6185). In contrast, the correlation for the entire clinical dataset (<i>n</i> = 420) yielded an R<sup>2</sup> of 0.4583. Comparison with existing literature indicates that the Inspire 6F© has a lower transfer capacity compared to larger membrane surface area oxygenators, as expected.ConclusionsThe newly developed oxygen transfer function provides a practical, evidence-based tool for the clinical setting. It allows perfusionists to adjust the FiO2 accurately to meet patient-specific oxygen requirements and maintain desired PaO2 levels. Implementing this function in routine practice can enhance standardization and improve patient safety by preventing hyperoxygenation.

Institutions

Publication Details

Journal
Perfusion
Published
2026-03-24
DOI
https://doi.org/10.1177/02676591261438976
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Oxygenator oxygen transfer performance during cardiopulmonary bypass of the Sorin ® Inspire 6F ©

Perfusion
Mechanical Circulatory Support Devices
article

Oxygenator oxygen transfer performance during cardiopulmonary bypass of the Sorin ® Inspire 6F ©

article en

Abstract

BackgroundOxygenator oxygen transfer performance is typically measured by manufacturers under standardized, static laboratory conditions. However, these parameters do not adequately reflect the dynamic environment of clinical cardiopulmonary bypass (CPB). To align with modern goal-directed perfusion standards - specifically the maintenance of physiological PaO2 levels to reduce oxidative stress - perfusionists require a reliable tool to titrate FiO2 based on real-time oxygen demand. This study aimed to develop a clinical oxygen transfer function for the Sorin® Inspire 6F© oxygenator.MethodsWe retrospectively analyzed 420 blood gas samples and hemodynamic parameters from 54 adult patients undergoing CPB with the Sorin® Inspire 6F©. Oxygen transfer (VO2) was calculated via the Fick principle. Using the methodology established by Hamilton et al., a validation subset (<i>n</i> = 57) of samples within a target PaO2 range of 150 ± 5 mmHg was used to derive a specific correlation function between FiO2 and VO2 via linear regression analysis.ResultsA significant correlation was observed between FiO2 and oxygen transfer. For the validation subset, the derived oxygen transfer function for the Sorin® Inspire 6F© was VO<sub>2</sub> = 284.34 x FiO2 (R<sup>2</sup> = 0.6185). In contrast, the correlation for the entire clinical dataset (<i>n</i> = 420) yielded an R<sup>2</sup> of 0.4583. Comparison with existing literature indicates that the Inspire 6F© has a lower transfer capacity compared to larger membrane surface area oxygenators, as expected.ConclusionsThe newly developed oxygen transfer function provides a practical, evidence-based tool for the clinical setting. It allows perfusionists to adjust the FiO2 accurately to meet patient-specific oxygen requirements and maintain desired PaO2 levels. Implementing this function in routine practice can enhance standardization and improve patient safety by preventing hyperoxygenation.

Perfusion
Heidelberg University (DE), University Hospital Heidelberg (DE), Hinckley and Bosworth Community Hospital (GB)
Openalex Percentile: Top 10%
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.