High-resolution in-beam video rate imaging with the ClearXCam for synchrotron beamlines
The ClearXCam detector is a new video rate imaging in-beam monitor based on a diamond sensor. Imaging with 2304 effective pixels is achieved by sequentially biasing one of 48 metal stripes on one side of a diamond sensor, while reading out the current from 48 stripes on the other side of the sensor. This system was characterized for real-time X-ray beam diagnostics at synchrotron beamline 17-BM at the National Synchrotron Light Source II. Significant results include: detection of in-beam structure via the imaging mode, beam focusing in one minute with real-time imaging feedback during a focusing event, linearity over five orders of magnitude, validation of a fast mode operating at 100 Hz, and sub-micron beam-positioning resolution. The system is now available commercially.
Institutions
- Brookhaven National Laboratory (US)
- Diamond Materials (United States) (US)
- Case Western Reserve University (US)
Publication Details
- Journal
- Journal of Synchrotron Radiation
- Published
- 2026-03-24
- DOI
- https://doi.org/10.1107/s1600577526001438
- Primary Topic
- Advanced X-ray Imaging Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00