Influence of Electrode–Tissue Contact Area on Parameter Sensitivity in Electrosurgical Monopolar Soft Coagulation: A Multiphysics Finite Element Study

Physics-based simulations are increasingly used to improve understanding of electrosurgical processes and to enable model-based estimation of tissue state when direct sensing is limited. The performance of such simulation-based virtual sensing approaches strongly depends on an accurate representation of the electrode-tissue interface. Despite its central role in electrical and thermal coupling, the influence of the electrode-tissue contact area has received limited attention in existing simulation studies. In this work, the influence of the electrode-tissue contact area on the sensitivity of key temperature-dependent tissue parameters was investigated for electrosurgical monopolar soft coagulation. Using a multiphysics finite element model under controlled boundary conditions, the sensitivity of maximum temperature development and necrotic tissue volume formation was analyzed with respect to varying contact areas and initial values of electrical conductivity, thermal conductivity, and effective heat capacity. The results demonstrate that parameter sensitivities are strongly contact-area-dependent. Electrical conductivity exhibits the most pronounced influence, particularly at larger contact areas, while thermal conductivity remains of minor relevance. In contrast, effective heat capacity significantly affects necrotic tissue volume formation, with increasing sensitivity for larger contact areas. These findings emphasize the importance of accurately accounting for electrode-tissue contact conditions in simulation-based analyses and clarify how contact-area-dependent sensitivities influence model-based tissue state estimation in electrosurgical coagulation.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2026-03-21
DOI
https://doi.org/10.3390/s26061975
Primary Topic
Thyroid and Parathyroid Surgery
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Influence of Electrode–Tissue Contact Area on Parameter Sensitivity in Electrosurgical Monopolar Soft Coagulation: A Multiphysics Finite Element Study

Christoph Busch, Knut Möeller, Stefan J. Rupitsch
Sensors
Thyroid and Parathyroid Surgery
article

Influence of Electrode–Tissue Contact Area on Parameter Sensitivity in Electrosurgical Monopolar Soft Coagulation: A Multiphysics Finite Element Study

Christoph Busch, Knut Möeller, Stefan J. Rupitsch
article en

Abstract

Physics-based simulations are increasingly used to improve understanding of electrosurgical processes and to enable model-based estimation of tissue state when direct sensing is limited. The performance of such simulation-based virtual sensing approaches strongly depends on an accurate representation of the electrode-tissue interface. Despite its central role in electrical and thermal coupling, the influence of the electrode-tissue contact area has received limited attention in existing simulation studies. In this work, the influence of the electrode-tissue contact area on the sensitivity of key temperature-dependent tissue parameters was investigated for electrosurgical monopolar soft coagulation. Using a multiphysics finite element model under controlled boundary conditions, the sensitivity of maximum temperature development and necrotic tissue volume formation was analyzed with respect to varying contact areas and initial values of electrical conductivity, thermal conductivity, and effective heat capacity. The results demonstrate that parameter sensitivities are strongly contact-area-dependent. Electrical conductivity exhibits the most pronounced influence, particularly at larger contact areas, while thermal conductivity remains of minor relevance. In contrast, effective heat capacity significantly affects necrotic tissue volume formation, with increasing sensitivity for larger contact areas. These findings emphasize the importance of accurately accounting for electrode-tissue contact conditions in simulation-based analyses and clarify how contact-area-dependent sensitivities influence model-based tissue state estimation in electrosurgical coagulation.

SensorsVol. 26(6)
University of Freiburg (DE), Furtwangen University (DE)
Openalex Percentile: Top 12%
Thyroid and Parathyroid Surgery
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.