Split-Sample Quality Assurance for the (1,3)-β-D-Glucan Assay

Background: (1,3)--D-Glucan (BDG) is an adjunctive biomarker for invasive fungal infections, but its analytical variability and lack of external quality assessment (EQA) programs limit assay reliability.This study evaluated BDG stability under practical storage conditions and explored the feasibility of a split-sample procedure as an alternative quality-assurance approach.Methods: The reliability of BDG assay results was assessed through three experiments addressing pre-analytical handling, storage stability, and interlaboratory comparability.To examine the effect of freeze-thaw stress, samples with low, moderate, and high BDG concentrations were repeatedly frozen and thawed.Short-term stability was then evaluated using pooled samples at near-cutoff, weak-positive, and strong-positive levels stored at different temperatures for up to 72 hours, with triplicate testing at predefined intervals.In addition, seven serum samples were aliquoted and analyzed in parallel at two institutions to assess interlaboratory agreement.Results: Repeated freeze-thaw cycles resulted in a marked reduction in BDG concentrations, whereas values remained largely stable for up to 72 hours under both refrigerated and frozen storage.Variability was the greatest near the assay cutoff.Interlaboratory comparison showed acceptable agreement in six of seven samples, with one discrepant result likely attributable to specimen-specific or handling-related factors.Conclusions: BDG assays remain stable under short-term refrigerated or frozen storage but are susceptible to repeated freeze-thaw cycles and variability near the cutoff range.The split-sample procedure offers a practical interim approach when EQA is unavailable and may serve as a foundation for standardized quality-assurance frameworks for BDG testing.

Authors

Institutions

Publication Details

Journal
Journal of Laboratory Medicine and Quality Assurance
Published
2026-03-24
DOI
https://doi.org/10.15263/jlmqa.2026.48.1.20
Primary Topic
Polysaccharides and Plant Cell Walls
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Split-Sample Quality Assurance for the (1,3)-β-D-Glucan Assay

Soo‐Kyung Kim, Youn Keong Cho, Heungsup Sung
Journal of Laboratory Medicine and Quality Assurance
Polysaccharides and Plant Cell Walls
article

Split-Sample Quality Assurance for the (1,3)-β-D-Glucan Assay

Soo‐Kyung Kim, Youn Keong Cho, Heungsup Sung
article en

Abstract

Background: (1,3)--D-Glucan (BDG) is an adjunctive biomarker for invasive fungal infections, but its analytical variability and lack of external quality assessment (EQA) programs limit assay reliability.This study evaluated BDG stability under practical storage conditions and explored the feasibility of a split-sample procedure as an alternative quality-assurance approach.Methods: The reliability of BDG assay results was assessed through three experiments addressing pre-analytical handling, storage stability, and interlaboratory comparability.To examine the effect of freeze-thaw stress, samples with low, moderate, and high BDG concentrations were repeatedly frozen and thawed.Short-term stability was then evaluated using pooled samples at near-cutoff, weak-positive, and strong-positive levels stored at different temperatures for up to 72 hours, with triplicate testing at predefined intervals.In addition, seven serum samples were aliquoted and analyzed in parallel at two institutions to assess interlaboratory agreement.Results: Repeated freeze-thaw cycles resulted in a marked reduction in BDG concentrations, whereas values remained largely stable for up to 72 hours under both refrigerated and frozen storage.Variability was the greatest near the assay cutoff.Interlaboratory comparison showed acceptable agreement in six of seven samples, with one discrepant result likely attributable to specimen-specific or handling-related factors.Conclusions: BDG assays remain stable under short-term refrigerated or frozen storage but are susceptible to repeated freeze-thaw cycles and variability near the cutoff range.The split-sample procedure offers a practical interim approach when EQA is unavailable and may serve as a foundation for standardized quality-assurance frameworks for BDG testing.

Journal of Laboratory Medicine and Quality AssuranceVol. 48(1)
Ulsan College (KR), Ewha Womans University (KR), Asan Medical Center (KR), University of Ulsan (KR)
Industry, innovation and infrastructure
Openalex Percentile: Top 12%
Polysaccharides and Plant Cell Walls
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.