Perioperative Validation of Two Handheld Glucometers in Dogs Under General Anesthesia: Analytical Robustness and Clinical Risk Assessment

Accurate perioperative glucose monitoring is essential in dogs undergoing general anesthesia, yet most validation studies of handheld glucometers have been performed under stable outpatient conditions. This prospective clinical validation study evaluated the analytical agreement, diagnostic performance, and ISO 15197 compliance of a human-calibrated (Accu-Chek) and a veterinary-specific (Centrivet GK) handheld glucometer compared with a laboratory spectrophotometric reference method in 34 anesthetized dogs (99 paired measurements per device). Linear mixed-effects modeling demonstrated significant method effects (<i>p</i> < 0.001), with the veterinary-specific device overestimating glucose concentrations relative to the reference method (β = 20.79 mg/dL; 95% CI: 8.08-33.50; <i>p</i> = 0.001), whereas the human-calibrated device did not differ significantly (β = 7.18 mg/dL; 95% CI: -5.53-19.89; <i>p</i> = 0.267). Bland-Altman analysis showed mean bias of 4.44 mg/dL (95% CI: 0.73-8.16) for the human-calibrated device and 22.72 mg/dL (95% CI: 18.22-27.21) for the veterinary-specific device. Passing-Bablok regression identified proportional bias only for the veterinary-specific device (slope 1.19; 95% CI: 1.01-1.34). ISO compliance was 69.7% and 39.4%, respectively. For hyperglycemia detection, AUC values were 0.9566 (95% CI: 0.8955-1.0000) and 0.9757 (95% CI: 0.9479-1.0000); for hypoglycemia, 0.8567 (95% CI: 0.7557-0.9578) and 0.7376 (95% CI: 0.6056-0.8697). In anesthetized dogs, the human-calibrated device demonstrated superior analytical robustness, whereas the veterinary-specific device showed greater bias and variability.

Authors

Institutions

Publication Details

Journal
Animals
Published
2026-03-23
DOI
https://doi.org/10.3390/ani16060993
Primary Topic
Hyperglycemia and glycemic control in critically ill and hospitalized patients
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Perioperative Validation of Two Handheld Glucometers in Dogs Under General Anesthesia: Analytical Robustness and Clinical Risk Assessment

Jorge U. Carmona, Valentina Hincapié, Catalina López
Animals
Hyperglycemia and glycemic control in critically ill and hospitalized patients
article

Perioperative Validation of Two Handheld Glucometers in Dogs Under General Anesthesia: Analytical Robustness and Clinical Risk Assessment

Jorge U. Carmona, Valentina Hincapié, Catalina López
article en

Abstract

Accurate perioperative glucose monitoring is essential in dogs undergoing general anesthesia, yet most validation studies of handheld glucometers have been performed under stable outpatient conditions. This prospective clinical validation study evaluated the analytical agreement, diagnostic performance, and ISO 15197 compliance of a human-calibrated (Accu-Chek) and a veterinary-specific (Centrivet GK) handheld glucometer compared with a laboratory spectrophotometric reference method in 34 anesthetized dogs (99 paired measurements per device). Linear mixed-effects modeling demonstrated significant method effects (<i>p</i> < 0.001), with the veterinary-specific device overestimating glucose concentrations relative to the reference method (β = 20.79 mg/dL; 95% CI: 8.08-33.50; <i>p</i> = 0.001), whereas the human-calibrated device did not differ significantly (β = 7.18 mg/dL; 95% CI: -5.53-19.89; <i>p</i> = 0.267). Bland-Altman analysis showed mean bias of 4.44 mg/dL (95% CI: 0.73-8.16) for the human-calibrated device and 22.72 mg/dL (95% CI: 18.22-27.21) for the veterinary-specific device. Passing-Bablok regression identified proportional bias only for the veterinary-specific device (slope 1.19; 95% CI: 1.01-1.34). ISO compliance was 69.7% and 39.4%, respectively. For hyperglycemia detection, AUC values were 0.9566 (95% CI: 0.8955-1.0000) and 0.9757 (95% CI: 0.9479-1.0000); for hypoglycemia, 0.8567 (95% CI: 0.7557-0.9578) and 0.7376 (95% CI: 0.6056-0.8697). In anesthetized dogs, the human-calibrated device demonstrated superior analytical robustness, whereas the veterinary-specific device showed greater bias and variability.

AnimalsVol. 16(6)
University of Caldas (CO)
Openalex Percentile: Top 15%
Hyperglycemia and glycemic control in critically ill and hospitalized patients
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.