Metrological Validation of Low-Cost DS18B20 Digital Temperature Sensors Using the TH-001 Procedure: Calibration Models, Uncertainty, and Reproducibility

This study presents the metrological validation of encapsulated DS18B20 digital temperature sensors. Eight units were tested, and seven were analysed (sensor 8 was excluded owing to a systematic failure). The evaluation was performed using a standard comparison calibration, where Tref was defined as the mean of two calibrated Pt-100 probes in a Julabo DYNEO DD 601F thermostatic bath, following the TH-001 procedure of the Spanish Centre of Metrology (CEM). Four validation tests were performed: Test 1 (E1, 20 to 75 °C), Test 2 (E2, 20 to 72 °C), and with an extended range, Test 3 (E3, −12 to 86 °C) and Test 4 (E4, −12 to 86 °C; repetition to assess reproducibility relative to E3), with 10 steady-state readings per setpoint. Erroneous readings were defined and removed (probe 3, Test 4), and set points without valid readings from probe 4 above 68 °C were excluded. Without data processing, the errors were consistent with the manufacturer’s stated ±0.5 °C, despite an inter-probe bias. Several correction models were evaluated (offset, affine linear, polynomial, and segmented); the probe-specific affine linear model provided the best overall compromise, reducing MAE (Mean Absolute Error) to 0.046 to 0.130 °C and RMSE (Root Mean Square Error) to 0.057 to 0.169 °C. The process uncertainty is dominated by the traceability of the Pt-100 probes and the effective nonuniformity of the isothermal volume, which limits the achievable accuracy. The results support the use of individually calibrated DS18B20 sensors for continuous monitoring, provided that the effective operating range is maintained.

Authors

Institutions

Publication Details

Journal
Metrology
Published
2026-03-23
DOI
https://doi.org/10.3390/metrology6010021
Primary Topic
Calibration and Measurement Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Metrological Validation of Low-Cost DS18B20 Digital Temperature Sensors Using the TH-001 Procedure: Calibration Models, Uncertainty, and Reproducibility

Jorge L. Costafreda Mustelier, Ana García Laso, Alfredo Marín Lázaro, Domingo Martín-Sánchez et al.
Metrology
Calibration and Measurement Techniques
article

Metrological Validation of Low-Cost DS18B20 Digital Temperature Sensors Using the TH-001 Procedure: Calibration Models, Uncertainty, and Reproducibility

Jorge L. Costafreda Mustelier, Ana García Laso, Alfredo Marín Lázaro, Domingo Martín-Sánchez, Javier Maroto Lorenzo, Leticia Presa Madrigal, Juan Antonio Rodríguez-Rama
article en

Abstract

This study presents the metrological validation of encapsulated DS18B20 digital temperature sensors. Eight units were tested, and seven were analysed (sensor 8 was excluded owing to a systematic failure). The evaluation was performed using a standard comparison calibration, where Tref was defined as the mean of two calibrated Pt-100 probes in a Julabo DYNEO DD 601F thermostatic bath, following the TH-001 procedure of the Spanish Centre of Metrology (CEM). Four validation tests were performed: Test 1 (E1, 20 to 75 °C), Test 2 (E2, 20 to 72 °C), and with an extended range, Test 3 (E3, −12 to 86 °C) and Test 4 (E4, −12 to 86 °C; repetition to assess reproducibility relative to E3), with 10 steady-state readings per setpoint. Erroneous readings were defined and removed (probe 3, Test 4), and set points without valid readings from probe 4 above 68 °C were excluded. Without data processing, the errors were consistent with the manufacturer’s stated ±0.5 °C, despite an inter-probe bias. Several correction models were evaluated (offset, affine linear, polynomial, and segmented); the probe-specific affine linear model provided the best overall compromise, reducing MAE (Mean Absolute Error) to 0.046 to 0.130 °C and RMSE (Root Mean Square Error) to 0.057 to 0.169 °C. The process uncertainty is dominated by the traceability of the Pt-100 probes and the effective nonuniformity of the isothermal volume, which limits the achievable accuracy. The results support the use of individually calibrated DS18B20 sensors for continuous monitoring, provided that the effective operating range is maintained.

MetrologyVol. 6(1)
Instituto Geológico y Minero de España (ES), European Telecommunications Standards Institute (FR), Universidad Politécnica de Madrid (ES)
Quality Education
Openalex Percentile: Top 11%
Calibration and Measurement Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.