A phantom study to achieve comparable 124I PET recovery and address scanner variability for dosimetry purposes in (re)differentiated thyroid cancer
Vendor-recommended <sup>124</sup>I acquisition and reconstruction leads to differences in quantification but can be compensated using a bias correction. Recovery correction is minimally affected by different BG-to-lesion ratios, suggesting that one RC curve is sufficient, simplifying <sup>124</sup>I calibration procedures in studies requiring <sup>124</sup>I quantification.