Comparison of Reproducibility and Repeatability of Corneal Topography Measured by ATLAS 500 and ATLAS 9000 in Average Patient Population

Pedro Tañá-Rivero, Paz Orts-Vila, Pedro Tañá-Sanz, Santiago Tañá-Sanz, Maria Ramos, Cristina Tello-Elordi Department of Ophthalmology, Oftalvist, Alicante, SpainCorrespondence: Pedro Tañá-Rivero, Department of Ophthalmology, Oftalvist, C/Angel Lozano 11, Alicante, 03001, Spain, Tel +34 965 14 15 00, Email [email protected]: To compare repeatability, reproducibility, and agreement of the Placido-based ATLAS 500 topographer (straight and swivel-out positions) with its predecessor, ATLAS 9000.Design: Prospective, single-center, post-market observational study.Methods: In this prospective, single-center post-market study involving 100 eyes with various corneal profiles, two operators obtained randomized, repeated measurements using the ATLAS 500 and ATLAS 9000 topographers. Parameters included corneal power, astigmatism, white-to-white distance, meridian powers, aberrations, and Placido coverage. Primary outcomes were repeatability and reproducibility; secondary analyses assessed coefficient of variation (CV), inter-device agreement, and keratoconus classification (Cohen’s κ).Results: The ATLAS 500 showed equal or superior repeatability and reproducibility compared to ATLAS 9000 across most parameters, with lower standard deviations (SDs) for central power (K1: 0.025 vs. 0.062 D; K2: 0.023 vs. 0.058 D) and astigmatism (0.015 vs. 0.018 D). The only exception was white-to-white distance, where ATLAS 9000 had lower variability (SD 0.032 mm vs. 0.066 mm), albeit with more missing data (41 vs. 10). The ATLAS 500 yielded lower coefficients of variation for corneal power and astigmatism (e.g. CV 0.058% vs. 0.146% for K1). Bland–Altman analysis showed minimal bias for K1/K2, with narrower limits in intra-device comparisons. Inter-device variability was greater for astigmatism and WTW, with ATLAS 500 measuring smaller WTW values. The swivel-out configuration increased Placido coverage by approximately 4%. Keratoconus classification showed moderate agreement (κ = 0.50) between devices.Conclusion: ATLAS 500 showed measurement precision comparable to or better than ATLAS 9000, with strong intra-device repeatability and broader corneal coverage in the swivel-out mode, confirming its suitability for routine corneal topography and refractive assessment.Keywords: corneal topography, placido disk, repeatability, reproducibility, bland–altman agreement, keratoconus, ATLAS 500, ATLAS 9000

Authors

Publication Details

Journal
Dove Medical Press (Taylor and Francis Group)
Published
2026-03-20
Primary Topic
Ophthalmology and Visual Impairment Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of Reproducibility and Repeatability of Corneal Topography Measured by ATLAS 500 and ATLAS 9000 in Average Patient Population

Cristina Tello-Elordi, Paz Orts-Vila, Santiago Tañá-Sanz, Pedro Tañá-Rivero et al.
Dove Medical Press (Taylor and Francis Group)
Ophthalmology and Visual Impairment Studies
article

Comparison of Reproducibility and Repeatability of Corneal Topography Measured by ATLAS 500 and ATLAS 9000 in Average Patient Population

Cristina Tello-Elordi, Paz Orts-Vila, Santiago Tañá-Sanz, Pedro Tañá-Rivero, Pedro Tañá-Sanz, Maria Ramos
article en

Abstract

Pedro Tañá-Rivero, Paz Orts-Vila, Pedro Tañá-Sanz, Santiago Tañá-Sanz, Maria Ramos, Cristina Tello-Elordi Department of Ophthalmology, Oftalvist, Alicante, SpainCorrespondence: Pedro Tañá-Rivero, Department of Ophthalmology, Oftalvist, C/Angel Lozano 11, Alicante, 03001, Spain, Tel +34 965 14 15 00, Email [email protected]: To compare repeatability, reproducibility, and agreement of the Placido-based ATLAS 500 topographer (straight and swivel-out positions) with its predecessor, ATLAS 9000.Design: Prospective, single-center, post-market observational study.Methods: In this prospective, single-center post-market study involving 100 eyes with various corneal profiles, two operators obtained randomized, repeated measurements using the ATLAS 500 and ATLAS 9000 topographers. Parameters included corneal power, astigmatism, white-to-white distance, meridian powers, aberrations, and Placido coverage. Primary outcomes were repeatability and reproducibility; secondary analyses assessed coefficient of variation (CV), inter-device agreement, and keratoconus classification (Cohen’s κ).Results: The ATLAS 500 showed equal or superior repeatability and reproducibility compared to ATLAS 9000 across most parameters, with lower standard deviations (SDs) for central power (K1: 0.025 vs. 0.062 D; K2: 0.023 vs. 0.058 D) and astigmatism (0.015 vs. 0.018 D). The only exception was white-to-white distance, where ATLAS 9000 had lower variability (SD 0.032 mm vs. 0.066 mm), albeit with more missing data (41 vs. 10). The ATLAS 500 yielded lower coefficients of variation for corneal power and astigmatism (e.g. CV 0.058% vs. 0.146% for K1). Bland–Altman analysis showed minimal bias for K1/K2, with narrower limits in intra-device comparisons. Inter-device variability was greater for astigmatism and WTW, with ATLAS 500 measuring smaller WTW values. The swivel-out configuration increased Placido coverage by approximately 4%. Keratoconus classification showed moderate agreement (κ = 0.50) between devices.Conclusion: ATLAS 500 showed measurement precision comparable to or better than ATLAS 9000, with strong intra-device repeatability and broader corneal coverage in the swivel-out mode, confirming its suitability for routine corneal topography and refractive assessment.Keywords: corneal topography, placido disk, repeatability, reproducibility, bland–altman agreement, keratoconus, ATLAS 500, ATLAS 9000

Dove Medical Press (Taylor and Francis Group)
Openalex Percentile: Top 16%
Ophthalmology and Visual Impairment Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.