Agreement of ocular biometry and evaluation of IOL calculation using a parameter‑hybrid strategy: a comparison of CASIA2, IOL Master 700, and OA2000 in cataract patients
The hybrid strategy introduces a hyperopia bias and reduces refractive prediction accuracy. Using a single, consistent device for preoperative measurements is recommended for routine cataract surgery. The hybrid approach may serve only as a contingency in complex cases.
Authors
- Shirong Chen (ORCID: https://orcid.org/0000-0002-0965-791X)
- Zhiqiang Guan (ORCID: https://orcid.org/0000-0002-6053-2717)
- Kunliang Qiu
- Weiwen Lin
- Siping Wu
- Xinyu Liu
- Yingyan Chen
Institutions
- Chinese University of Hong Kong (HK)
- Shantou University (CN)
Publication Details
- Journal
- BMC Ophthalmology
- Published
- 2026-03-17
- DOI
- https://doi.org/10.1186/s12886-026-04736-6
- Primary Topic
- Ophthalmology and Visual Impairment Studies
- Type
- article
- Field-Weighted Citation Impact
- 0.00