Assessment of radiation doses in computed tomography scans using simulation models versus experimental verification and cancer risk estimation in pediatric phantom

Abstract Objective This study aims to develop a methodology for estimating the absorbed dose in various organs of a pediatric phantom during computed tomography (CT) scans. The approach combines Monte Carlo simulations with experimental validation to ensure the accuracy of the numerical results. Methods This study employed thermoluminescent dosimeters (TLDs) and GEANT4/GATE-based Monte Carlo simulations to estimate and validate organ dose measurements in a pediatric phantom. The phantom was irradiated utilizing a Siemens CT scanner at appropriate parameters (120 kVp and 60 mA). We calculated the dose at identical positions within the phantom using Monte Carlo simulations and subsequently compared the dose results. Results The absorbed doses recorded by the TLDs for the left eye, right eye, thyroid, left lung, right lung, left pelvic, right pelvic, and testes were 18.93 ± 1.02, 16.47 ± 0.86, 28.37 ± 1.63, 16.30 ± 1.06, 15.21 ± 1.26, 14.33 ± 0.01, 14.10 ± 1.6, and 17.91 ± 0.16 mGy, respectively. The calculated absorbed doses in the respective organs using GEANT4/GATE were: 16.96 ± 1.99, 16.75 ± 1.45, 29.61 ± 0.77, 16.92 ± 0.08, 16.26 ± 0.39, 15.07 ± 0.01, 14.91 ± 0.75, and 17.94 ± 0.041 mGy. The results indicate a strong concordance between the two methods, with percentage differences of 11.6%, 1.7%, 4.4%, 3.77%, 6.9%, 5.14%, 5.74%, and 0.17%, respectively. Conclusion The strong agreement between simulation results and experimental measurements enhances the ability to estimate absorbed doses in radiosensitive organs during CT scans. Additionally, the study utilizes phantom dose data to evaluate the risk of secondary cancers in susceptible organs, highlighting a relationship between lifetime risk and excess absolute risk for both males and females.

Authors

Institutions

Publication Details

Journal
The Egyptian Journal of Radiology and Nuclear Medicine
Published
2026-03-17
DOI
https://doi.org/10.1186/s43055-026-01715-1
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessment of radiation doses in computed tomography scans using simulation models versus experimental verification and cancer risk estimation in pediatric phantom

Rasha A. Elawady, Abdelrahman Mosallam, Raghda A. El Gendy, Shireen A. Abdelbadee et al.
The Egyptian Journal of Radiology and Nuclear Medicine
Radiation Dose and Imaging
article

Assessment of radiation doses in computed tomography scans using simulation models versus experimental verification and cancer risk estimation in pediatric phantom

Rasha A. Elawady, Abdelrahman Mosallam, Raghda A. El Gendy, Shireen A. Abdelbadee, Ehab M. Attalla
article en

Abstract

Abstract Objective This study aims to develop a methodology for estimating the absorbed dose in various organs of a pediatric phantom during computed tomography (CT) scans. The approach combines Monte Carlo simulations with experimental validation to ensure the accuracy of the numerical results. Methods This study employed thermoluminescent dosimeters (TLDs) and GEANT4/GATE-based Monte Carlo simulations to estimate and validate organ dose measurements in a pediatric phantom. The phantom was irradiated utilizing a Siemens CT scanner at appropriate parameters (120 kVp and 60 mA). We calculated the dose at identical positions within the phantom using Monte Carlo simulations and subsequently compared the dose results. Results The absorbed doses recorded by the TLDs for the left eye, right eye, thyroid, left lung, right lung, left pelvic, right pelvic, and testes were 18.93 ± 1.02, 16.47 ± 0.86, 28.37 ± 1.63, 16.30 ± 1.06, 15.21 ± 1.26, 14.33 ± 0.01, 14.10 ± 1.6, and 17.91 ± 0.16 mGy, respectively. The calculated absorbed doses in the respective organs using GEANT4/GATE were: 16.96 ± 1.99, 16.75 ± 1.45, 29.61 ± 0.77, 16.92 ± 0.08, 16.26 ± 0.39, 15.07 ± 0.01, 14.91 ± 0.75, and 17.94 ± 0.041 mGy. The results indicate a strong concordance between the two methods, with percentage differences of 11.6%, 1.7%, 4.4%, 3.77%, 6.9%, 5.14%, 5.74%, and 0.17%, respectively. Conclusion The strong agreement between simulation results and experimental measurements enhances the ability to estimate absorbed doses in radiosensitive organs during CT scans. Additionally, the study utilizes phantom dose data to evaluate the risk of secondary cancers in susceptible organs, highlighting a relationship between lifetime risk and excess absolute risk for both males and females.

The Egyptian Journal of Radiology and Nuclear MedicineVol. 57(1)
Cairo University (EG), Fayoum University (EG), Alexandria University (EG)
Good health and well-being
Openalex Percentile: Top 20%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.