Can the current radiation dose, in chest tomosynthesis, be reduced with retained image quality? A study in the context of lung cancer screening programs
This study has not shown any reduction in image quality from reducing the radiation dose by 30%, compared with the vendor-recommended protocol, which might indicate that such a reduction could be used in clinical practice.