Development and Validation of Reverse Phase High-Performance Liquid Chromatographic Method Enabling Simultaneous Estimation of Polmacoxib and its Process-Related Impurities: Exploring Degradation Pathways and Elucidating Degradant Structures via LC–M

A simple, sensitive, and selective reverse phase high-performance liquid chromatography (HPLC) method was developed and validated for the estimation of Polmacoxib (POL) and its process-related impurities. Chromatographic separation was achieved on a Kromasil C18 column (250 mm x 4.6 mm, 5 μm) using gradient elution with mobile phase containing 10 mM ammonium acetate buffer and acetonitrile in the initial ratio of 90:10 (% v/v). Chromatographic parameters were optimized: flow rate of 1.2 mL/min, UV detection at 232 nm, and injection volume of 10 μL. The developed method was validated as per International Council for Harmonization Q2 (R1) guidelines. For the stability study, the drug was subjected to various stress conditions. The results showed that was sensitive to the alkaline and degraded. Under alkaline condition, impurity-I and II were detected at retention time of 2.61 and 21.09 min, respectively. To identify the structures of the detected impurities, were initially isolated using a preparative HPLC and then collected fractions were injected into liquid chromatography with tandem mass spectrometry for structural identification. Further structure elucidation was performed using 1D-NMR (1H and 13C) and 2D-NMR (1H-1H correlation spectroscopy and 1H-13C heteronuclear single quantum coherence), and explore the degradation pathway. The proposed method is applicable for assay estimation of POL in dosage form and is suitable for quality control analysis.

Authors

Institutions

Publication Details

Journal
Journal of Chromatographic Science
Published
2026-02-13
DOI
https://doi.org/10.1093/chromsci/bmag006
Primary Topic
Analytical Methods in Pharmaceuticals
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development and Validation of Reverse Phase High-Performance Liquid Chromatographic Method Enabling Simultaneous Estimation of Polmacoxib and its Process-Related Impurities: Exploring Degradation Pathways and Elucidating Degradant Structures via LC–M

Bhavin Dhaduk, Hitesh Thumbar
Journal of Chromatographic Science
Analytical Methods in Pharmaceuticals
article

Development and Validation of Reverse Phase High-Performance Liquid Chromatographic Method Enabling Simultaneous Estimation of Polmacoxib and its Process-Related Impurities: Exploring Degradation Pathways and Elucidating Degradant Structures via LC–M

Bhavin Dhaduk, Hitesh Thumbar
article en

Abstract

A simple, sensitive, and selective reverse phase high-performance liquid chromatography (HPLC) method was developed and validated for the estimation of Polmacoxib (POL) and its process-related impurities. Chromatographic separation was achieved on a Kromasil C18 column (250 mm x 4.6 mm, 5 μm) using gradient elution with mobile phase containing 10 mM ammonium acetate buffer and acetonitrile in the initial ratio of 90:10 (% v/v). Chromatographic parameters were optimized: flow rate of 1.2 mL/min, UV detection at 232 nm, and injection volume of 10 μL. The developed method was validated as per International Council for Harmonization Q2 (R1) guidelines. For the stability study, the drug was subjected to various stress conditions. The results showed that was sensitive to the alkaline and degraded. Under alkaline condition, impurity-I and II were detected at retention time of 2.61 and 21.09 min, respectively. To identify the structures of the detected impurities, were initially isolated using a preparative HPLC and then collected fractions were injected into liquid chromatography with tandem mass spectrometry for structural identification. Further structure elucidation was performed using 1D-NMR (1H and 13C) and 2D-NMR (1H-1H correlation spectroscopy and 1H-13C heteronuclear single quantum coherence), and explore the degradation pathway. The proposed method is applicable for assay estimation of POL in dosage form and is suitable for quality control analysis.

Journal of Chromatographic ScienceVol. 64(3)
RK University (IN), Atmiya University (IN)
Partnerships for the goals
Openalex Percentile: Top 20%
Analytical Methods in Pharmaceuticals
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Development and Validation of Reverse Phase High-Performance Liquid Chromatographic Method Enabling Simultaneous Estimation of Polmacoxib and its Process-Related Impurities: Exploring Degradation Pathways and Elucidating Degradant Structures via LC–M — Bhavin Dhaduk, Hitesh Thumbar · Journal of Chromatographic Science (2026) | TGRS Research Map | TGRS