Oxygen Detection in Dermal Interstitial Fluid with Microneedles Based on Platinum Nanoparticles

Herein, we have developed a microneedle-based oxygen sensor wearable patch (O<sub>2</sub> MN) designed for real-time, minimally invasive detection of dermal ISF oxygen levels. This work pioneers the use of platinum nanoparticles as the sensing element for oxygen detection within a microneedle platform, enabling high electrochemical sensitivity at a mild applied potential. The wearable patch integrates the nanoparticle-based microneedle as the working electrode and a Ag/AgCl ink-modified microneedle as the counter/reference electrode, operating at an applied potential of -0.3 V to detect a wide range of oxygen (10-254 μM) with high sensitivity (-0.51 nA μM<sup>-1</sup>) and a rapid response time (9 s). The analytical assessment revealed excellent performance, including high reversibility (RSD = 3.4%), repeatability (RSD = 5.2%), medium-term stability (1.9% deviation over 90 min), acceptable reproducibility (6.6%), and mechanical reliability during skin insertion. Notably, to the best of our knowledge, this study represents the first comprehensive evaluation of the performance of an oxygen microneedle sensor, including systematic in vitro, on-body, and in vivo experiments. The sensor's accuracy was validated with a Pearson correlation coefficient of 0.99 for in vitro tests, and on-body oxygen measurements in euthanized rats highlighted the reliability of the O<sub>2</sub> MN patch for dermal ISF oxygen detection. In vivo studies further revealed a strong positive correlation (Pearson correlation coefficient = 0.83) between dermal ISF and blood oxygen levels, underscoring the O<sub>2</sub> MN patch's potential for biomedical applications through real-time, minimally invasive monitoring.

Authors

Institutions

Publication Details

Journal
ACS Sensors
Published
2026-03-16
DOI
https://doi.org/10.1021/acssensors.5c04533
Primary Topic
Advancements in Transdermal Drug Delivery
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Oxygen Detection in Dermal Interstitial Fluid with Microneedles Based on Platinum Nanoparticles

Asa K Konradsson-Geuken, Q. Wang, D.P. Rojas, Gastón A. Crespo et al.
ACS Sensors
Advancements in Transdermal Drug Delivery
article

Oxygen Detection in Dermal Interstitial Fluid with Microneedles Based on Platinum Nanoparticles

Asa K Konradsson-Geuken, Q. Wang, D.P. Rojas, Gastón A. Crespo, Qikun Wei, Ruben Zapata-Pérez, Agueda Molinero-Fernandez, María Cuartero
article en

Abstract

Herein, we have developed a microneedle-based oxygen sensor wearable patch (O<sub>2</sub> MN) designed for real-time, minimally invasive detection of dermal ISF oxygen levels. This work pioneers the use of platinum nanoparticles as the sensing element for oxygen detection within a microneedle platform, enabling high electrochemical sensitivity at a mild applied potential. The wearable patch integrates the nanoparticle-based microneedle as the working electrode and a Ag/AgCl ink-modified microneedle as the counter/reference electrode, operating at an applied potential of -0.3 V to detect a wide range of oxygen (10-254 μM) with high sensitivity (-0.51 nA μM<sup>-1</sup>) and a rapid response time (9 s). The analytical assessment revealed excellent performance, including high reversibility (RSD = 3.4%), repeatability (RSD = 5.2%), medium-term stability (1.9% deviation over 90 min), acceptable reproducibility (6.6%), and mechanical reliability during skin insertion. Notably, to the best of our knowledge, this study represents the first comprehensive evaluation of the performance of an oxygen microneedle sensor, including systematic in vitro, on-body, and in vivo experiments. The sensor's accuracy was validated with a Pearson correlation coefficient of 0.99 for in vitro tests, and on-body oxygen measurements in euthanized rats highlighted the reliability of the O<sub>2</sub> MN patch for dermal ISF oxygen detection. In vivo studies further revealed a strong positive correlation (Pearson correlation coefficient = 0.83) between dermal ISF and blood oxygen levels, underscoring the O<sub>2</sub> MN patch's potential for biomedical applications through real-time, minimally invasive monitoring.

ACS Sensors
Uppsala University (SE), American Society of Addiction Medicine (US), Universidad Católica de Murcia (ES), KTH Royal Institute of Technology (SE)
Openalex Percentile: Top 19%
Advancements in Transdermal Drug Delivery
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.