Tendon-based power transmission mechanism and collocated controller with application to position tracking of autonomous surgical robots

Due to compliance and significant friction in the drive mechanism, tendon-based minimally invasive surgical robot systems exhibit highly nonlinear behavior and low position control accuracy. In this study, a novel power transmission mechanism incorporating both tendon position and tension feedback was developed. The mechanism is controlled by a corresponding collocated controller based on position sensing and a cable tension sensing method, which enables compensation for nonlinearities in the motion transmission of the tendon-based surgical robot system. The collocated controller was implemented and experimentally validated on a surgical instrument utilizing the power transmission mechanism. Experimental results demonstrate that position tracking in the motion transmission of the tendon-driven instrument can be improved by 80% using the proposed mechanism and controller. This advancement in robotic surgery holds significant potential to enhance the precision and effectiveness of surgical procedures.

Authors

Institutions

Publication Details

Journal
Transactions of the Institute of Measurement and Control
Published
2026-03-15
DOI
https://doi.org/10.1177/01423312261425524
Primary Topic
Soft Robotics and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Tendon-based power transmission mechanism and collocated controller with application to position tracking of autonomous surgical robots

Evgeni Magid, Hongbing Li, Linlong Li, Yubin Lan et al.
Transactions of the Institute of Measurement and Control
Soft Robotics and Applications
article

Tendon-based power transmission mechanism and collocated controller with application to position tracking of autonomous surgical robots

Evgeni Magid, Hongbing Li, Linlong Li, Yubin Lan, Qilian Bao, Rui Wang
article en

Abstract

Due to compliance and significant friction in the drive mechanism, tendon-based minimally invasive surgical robot systems exhibit highly nonlinear behavior and low position control accuracy. In this study, a novel power transmission mechanism incorporating both tendon position and tension feedback was developed. The mechanism is controlled by a corresponding collocated controller based on position sensing and a cable tension sensing method, which enables compensation for nonlinearities in the motion transmission of the tendon-based surgical robot system. The collocated controller was implemented and experimentally validated on a surgical instrument utilizing the power transmission mechanism. Experimental results demonstrate that position tracking in the motion transmission of the tendon-driven instrument can be improved by 80% using the proposed mechanism and controller. This advancement in robotic surgery holds significant potential to enhance the precision and effectiveness of surgical procedures.

Transactions of the Institute of Measurement and Control
Shanghai Jiao Tong University (CN), Kazan Federal University (RU), Inner Mongolia People's Hospital (CN)
Openalex Percentile: Top 21%
Soft Robotics and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.