Patient-Specific Guide using Radiopaque Alignment for Minimally Invasive Sacroiliac Joint Fusion

1. Introduction <br/>Severe pelvic instability accounts for up to 30% of low back pain. When conservative treatment fails, sacroiliac joint fusion (SIJF) may be indicated to reduce pain by placing three implants inside the joint. SIJF is commonly performed using 2D fluoroscopic guidance. Due to anatomical variability, poor visibility during intraoperative imaging, and the absence of 3D spatial information, placing the implants without damaging neural structures is challenging. <br/>2. Objectives<br/>To improve patient outcomes, a workflow with a patient-specific guide (PSG) using alignment tools has been developed. A human cadaveric test was performed to assess whether the workflow can be safely implemented in clinical practice.<br/>3. Methods<br/>During typical PSG surgeries, a guide is seated onto the bone based on bony landmarks, enabling guided cutting or drilling according to a prespecified plan. Since SIJF is a minimally invasive procedure, bone seating is not possible. Therefore, a novel alignment tool was incorporated in the PSG, allowing it to be aligned using 2D fluoroscopy. The PSG was applied to five specimens, with both SI joints fused, comprising ten surgeries. Implant malposition complications and implant placement accuracy were determined.<br/>4. Results<br/>Thirty implants were placed with two malposition complications. Implant placement accuracy showed a positional deviation of 8.4 ± 3.5 mm and an angular deviation of 3.2 ± 1.9°.<br/>5. Conclusion<br/>Using radiopaque alignment tools in a PSG is a novel approach for guide alignment. However, the results were inferior to a previously developed technique involving virtual surgical planning. More research is necessary to optimize the workflow.<br/>

Authors

Institutions

Publication Details

Journal
University of Twente Research Information
Published
2025-04-17
Primary Topic
Spine and Intervertebral Disc Pathology
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Patient-Specific Guide using Radiopaque Alignment for Minimally Invasive Sacroiliac Joint Fusion

Jorm M. Nellensteijn, Nick Kampkuiper, MA Koenrades, E.E.G.; id_orcid 0000-0002-7198-6810 Hekman et al.
University of Twente Research Information
Spine and Intervertebral Disc Pathology
article

Patient-Specific Guide using Radiopaque Alignment for Minimally Invasive Sacroiliac Joint Fusion

Jorm M. Nellensteijn, Nick Kampkuiper, MA Koenrades, E.E.G.; id_orcid 0000-0002-7198-6810 Hekman, Gabriëlle; id_orcid 0000-0003-4372-6584 Tuijthof, Femke F.; id_orcid 0000-0003-0663-7873 Schröder
article en

Abstract

1. Introduction <br/>Severe pelvic instability accounts for up to 30% of low back pain. When conservative treatment fails, sacroiliac joint fusion (SIJF) may be indicated to reduce pain by placing three implants inside the joint. SIJF is commonly performed using 2D fluoroscopic guidance. Due to anatomical variability, poor visibility during intraoperative imaging, and the absence of 3D spatial information, placing the implants without damaging neural structures is challenging. <br/>2. Objectives<br/>To improve patient outcomes, a workflow with a patient-specific guide (PSG) using alignment tools has been developed. A human cadaveric test was performed to assess whether the workflow can be safely implemented in clinical practice.<br/>3. Methods<br/>During typical PSG surgeries, a guide is seated onto the bone based on bony landmarks, enabling guided cutting or drilling according to a prespecified plan. Since SIJF is a minimally invasive procedure, bone seating is not possible. Therefore, a novel alignment tool was incorporated in the PSG, allowing it to be aligned using 2D fluoroscopy. The PSG was applied to five specimens, with both SI joints fused, comprising ten surgeries. Implant malposition complications and implant placement accuracy were determined.<br/>4. Results<br/>Thirty implants were placed with two malposition complications. Implant placement accuracy showed a positional deviation of 8.4 ± 3.5 mm and an angular deviation of 3.2 ± 1.9°.<br/>5. Conclusion<br/>Using radiopaque alignment tools in a PSG is a novel approach for guide alignment. However, the results were inferior to a previously developed technique involving virtual surgical planning. More research is necessary to optimize the workflow.<br/>

University of Twente Research Information
The Medical Device (United Kingdom) (GB)
Openalex Percentile: Top 24%
Spine and Intervertebral Disc Pathology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.