In vitro VV-ECMO mock circuit to quantify oxygen transfer across ARDS and ECMO weaning stages

A continuous dual-loop mock circuit for venovenous extracorporeal membrane oxygenation (VV-ECMO) was developed to quantify oxygenation contributions across simulated pathophysiological (acute respiratory distress syndrome, ARDS) and therapy stages (ECMO weaning). ARDS stages (mild to severe) were calibrated for simulated patients using Berlin-definition targets and model-derived setpoints. The calculation of relative partitioning, oxygen transfer rate, and gas exchange ratio was enabled by four-site blood gas measurement. The results indicate the presence of distinct plateaus and a consistent shift of oxygenation from ECMO to lung compartment during ECMO weaning. The setup provides a foundation for future gas exchange models and weaning strategies.

Authors

Institutions

Publication Details

Journal
Infinite Science GmbH
Published
2025-12-17
DOI
https://doi.org/10.18416/automed.2026.2535
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

In vitro VV-ECMO mock circuit to quantify oxygen transfer across ARDS and ECMO weaning stages

Andreas Döcke, Susanne Kromnik, Hagen Malberg, Luisa Neubauer
Infinite Science GmbH
Mechanical Circulatory Support Devices
article

In vitro VV-ECMO mock circuit to quantify oxygen transfer across ARDS and ECMO weaning stages

Andreas Döcke, Susanne Kromnik, Hagen Malberg, Luisa Neubauer
article en

Abstract

A continuous dual-loop mock circuit for venovenous extracorporeal membrane oxygenation (VV-ECMO) was developed to quantify oxygenation contributions across simulated pathophysiological (acute respiratory distress syndrome, ARDS) and therapy stages (ECMO weaning). ARDS stages (mild to severe) were calibrated for simulated patients using Berlin-definition targets and model-derived setpoints. The calculation of relative partitioning, oxygen transfer rate, and gas exchange ratio was enabled by four-site blood gas measurement. The results indicate the presence of distinct plateaus and a consistent shift of oxygenation from ECMO to lung compartment during ECMO weaning. The setup provides a foundation for future gas exchange models and weaning strategies.

Infinite Science GmbH
Technische Universität Dresden (DE)
Openalex Percentile: Top 37%
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.