GFAP/UCH-L1 Assay for mTBI: Preanalytical Factors and Levels in Healthy Individuals—A Study on Danish Adults
The analytical precision and storage conditions of GFAP/UCH-L1 were satisfactory. Serum and plasma can be used for measurement of GFAP, whereas UCH-L1 measurement in plasma is not advised until further evaluations are conducted. The manufacturer's upper reference limits may not be suitable for our population. Thorough research on cutoff values is necessary.