Towards data-driven predictive temperature control for retinal laser treatments

Laser-induced coagulation is a common treatment of retinal diseases. Manual dosing is prone to error and can lead to extended damage of the neural retina. We elaborated a novel data-driven predictive temperature control (DPC). To identify controller hyperparameters with minimal deviation from the target temperature extensive simulation was carried out, based on 206 measured data sets, each applied pulse energy and induced temperature rise. This resulted in a median and mean deviation from the target temperature of 1.6 °C and 2.6 °C (95 % CI = 2.3 - 3.1 °C), respectively. In addition, we developed a detection algorithm for eye movements (saccades), optimized with Bayesian Optimization that achieved detection sensitivity of 92 % and specificity of 100 %. The experience with Bayesian Optimization shall be used to further optimize the DPC and to evaluate clinical data.

Authors

Institutions

Publication Details

Journal
Infinite Science GmbH
Published
2025-12-04
DOI
https://doi.org/10.18416/automed.2026.2467
Primary Topic
Retinal Imaging and Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Towards data-driven predictive temperature control for retinal laser treatments

Hossam S. Abbas, Henrik Volkens, Andra-Karina Poetschki, Ralf Brinkmann et al.
Infinite Science GmbH
Retinal Imaging and Analysis
article

Towards data-driven predictive temperature control for retinal laser treatments

Hossam S. Abbas, Henrik Volkens, Andra-Karina Poetschki, Ralf Brinkmann, Marietta Sophie Huch
article en

Abstract

Laser-induced coagulation is a common treatment of retinal diseases. Manual dosing is prone to error and can lead to extended damage of the neural retina. We elaborated a novel data-driven predictive temperature control (DPC). To identify controller hyperparameters with minimal deviation from the target temperature extensive simulation was carried out, based on 206 measured data sets, each applied pulse energy and induced temperature rise. This resulted in a median and mean deviation from the target temperature of 1.6 °C and 2.6 °C (95 % CI = 2.3 - 3.1 °C), respectively. In addition, we developed a detection algorithm for eye movements (saccades), optimized with Bayesian Optimization that achieved detection sensitivity of 92 % and specificity of 100 %. The experience with Bayesian Optimization shall be used to further optimize the DPC and to evaluate clinical data.

Infinite Science GmbH
Medizinisches Laserzentrum Lübeck (Germany) (DE), University of Lübeck (DE)
Openalex Percentile: Top 25%
Retinal Imaging and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.