Assessing the influence of kernel selection on chest computed tomography image quality across varying dose levels using TrueFidelity reconstruction

Deep learning image reconstruction (DLIR) utilizes neural networks to generate high-quality computed tomography (CT) images. One commercially available DLIR software is TrueFidelity from GE Healthcare. The Standard kernel was the only available reconstruction kernel previously, but recently other kernels, including the Lung kernel, have been introduced by GE. This study aimed to evaluate the image quality of chest CT scans acquired at full-dose (FD, 2.5 mSv) and ultra-low-dose (ULD, 0.05 mSv) when reconstructed using TrueFidelity with both Standard and Lung kernels. Twenty-five patients underwent chest CT scans at Sahlgrenska University Hospital. The images were reconstructed and then evaluated by four radiologists in two different studies, one including ULD CT axial images and the other one the FDCT. Visual Grading Characteristics (VGC) analysis was applied, using the Standard kernel as reference and the area under the VGC curve (AUCVGC) for comparison. At FD, the Standard kernel yielded better results regarding the visualization of six structures and the general image quality. However, in ULD scans, the differences between kernels were not statistically significant. The FD images were mostly rated as acceptable, while ULD images were often rated as probably acceptable or unacceptable, especially for emphysema assessment. Overall, TrueFidelity seems to perform better with the Standard kernel than with the Lung kernel in FD protocols, but no reliable conclusions can be drawn for the ULD protocol.

Authors

Institutions

Publication Details

Journal
Radiation Protection Dosimetry
Published
2025-11-03
DOI
https://doi.org/10.1093/rpd/ncaf143
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessing the influence of kernel selection on chest computed tomography image quality across varying dose levels using TrueFidelity reconstruction

Micael Oliveira Diniz, Angelica Svalkvist, Walter Cifuentes Ramirez, Rauni Rossi Norrlund et al.
Radiation Protection Dosimetry
Radiation Dose and Imaging
article

Assessing the influence of kernel selection on chest computed tomography image quality across varying dose levels using TrueFidelity reconstruction

Micael Oliveira Diniz, Angelica Svalkvist, Walter Cifuentes Ramirez, Rauni Rossi Norrlund, Magnus Båth, Eleftheria Gianko, Åse A Johnsson
article en

Abstract

Deep learning image reconstruction (DLIR) utilizes neural networks to generate high-quality computed tomography (CT) images. One commercially available DLIR software is TrueFidelity from GE Healthcare. The Standard kernel was the only available reconstruction kernel previously, but recently other kernels, including the Lung kernel, have been introduced by GE. This study aimed to evaluate the image quality of chest CT scans acquired at full-dose (FD, 2.5 mSv) and ultra-low-dose (ULD, 0.05 mSv) when reconstructed using TrueFidelity with both Standard and Lung kernels. Twenty-five patients underwent chest CT scans at Sahlgrenska University Hospital. The images were reconstructed and then evaluated by four radiologists in two different studies, one including ULD CT axial images and the other one the FDCT. Visual Grading Characteristics (VGC) analysis was applied, using the Standard kernel as reference and the area under the VGC curve (AUCVGC) for comparison. At FD, the Standard kernel yielded better results regarding the visualization of six structures and the general image quality. However, in ULD scans, the differences between kernels were not statistically significant. The FD images were mostly rated as acceptable, while ULD images were often rated as probably acceptable or unacceptable, especially for emphysema assessment. Overall, TrueFidelity seems to perform better with the Standard kernel than with the Lung kernel in FD protocols, but no reliable conclusions can be drawn for the ULD protocol.

Radiation Protection DosimetryVol. 202(3–4)
Sahlgrenska University Hospital (SE), Region Västra Götaland (SE), University of Gothenburg (SE)
Openalex Percentile: Top 25%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.