Backscatter from radiation protective drapes in fluoroscopically guided interventions – a patient dose to consider?

Radiation protective drapes are sometimes placed on the patient during fluoroscopically guided procedures to reduce scattered radiation to medical personnel. However, there are concerns that these drapes may increase patient radiation dose due to backscattered radiation. In this study, thermoluminescent dosimeters (TLDs) were used to quantify backscattered radiation from three protective drapes, on top of and at different depths, in a polymethyl methacrylate (PMMA phantom). Monte Carlo simulations were also performed to theoretically quantify backscattered radiation from lead. The results showed that the backscattered radiation varied with the material of the drape and was less than a factor of 0.001 of the radiation dose from the primary radiation beam (including backscatter from the PMMA phantom). This minor increase in patient radiation dose does not need to be considered if the use of such drapes has been justified from an occupational radiation protection perspective.

Authors

Institutions

Publication Details

Journal
Radiation Protection Dosimetry
Published
2026-01-22
DOI
https://doi.org/10.1093/rpd/ncaf188
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Backscatter from radiation protective drapes in fluoroscopically guided interventions – a patient dose to consider?

Jens Hemmingsson, Charlotta Lundh, Pernilla Jonasson, Maria Larsson
Radiation Protection Dosimetry
Radiation Dose and Imaging
article

Backscatter from radiation protective drapes in fluoroscopically guided interventions – a patient dose to consider?

Jens Hemmingsson, Charlotta Lundh, Pernilla Jonasson, Maria Larsson
article en

Abstract

Radiation protective drapes are sometimes placed on the patient during fluoroscopically guided procedures to reduce scattered radiation to medical personnel. However, there are concerns that these drapes may increase patient radiation dose due to backscattered radiation. In this study, thermoluminescent dosimeters (TLDs) were used to quantify backscattered radiation from three protective drapes, on top of and at different depths, in a polymethyl methacrylate (PMMA phantom). Monte Carlo simulations were also performed to theoretically quantify backscattered radiation from lead. The results showed that the backscattered radiation varied with the material of the drape and was less than a factor of 0.001 of the radiation dose from the primary radiation beam (including backscatter from the PMMA phantom). This minor increase in patient radiation dose does not need to be considered if the use of such drapes has been justified from an occupational radiation protection perspective.

Radiation Protection DosimetryVol. 202(3–4)
Sahlgrenska University Hospital (SE), University of Gothenburg (SE)
Openalex Percentile: Top 23%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.