Pilot assessment of point-of-care differential white blood cell count analysers in schoolchildren with schistosomiasis
Public health monitoring of schistosomiasis needs appropriate field-based technology, so to expand this portfolio, we conducted a first pilot assessment of point-of-care differential white blood cell count analysers. In total, 120 Malawian primary school children were examined with six key haematological parameters recorded. A notable proportion had abnormal differential white blood cell profiles with eosinophilia (48.3%) positively associated with schistosomiasis (OR 3.2, 95% CI 1.3 to 7.7). Whilst costs of equipment and associated tests were non-trivial, both machine performance and haematological information gained were satisfactory. To improve monitoring of schistosomiasis-associated pathologies, the inclusion of differential white blood cell count technology screening appears sensible.
Authors
- Peter Makaula (ORCID: https://orcid.org/0000-0002-1198-1655)
- Janelisa Musaya (ORCID: https://orcid.org/0000-0003-2928-0133)
- Sekeleghe A. Kayuni (ORCID: https://orcid.org/0000-0002-6948-0087)
- Sarah Rollason (ORCID: https://orcid.org/0000-0003-4330-7099)
- J Russell Stothard
- Jordan White
Institutions
- Liverpool School of Tropical Medicine (GB)
- Queen Elizabeth Central Hospital (MW)
- Cardiff University (GB)
Publication Details
- Journal
- Transactions of the Royal Society of Tropical Medicine and Hygiene
- Published
- 2026-02-21
- DOI
- https://doi.org/10.1093/trstmh/trag025
- Primary Topic
- Parasites and Host Interactions
- Type
- article
- Field-Weighted Citation Impact
- 0.00