Low-Voltage Blood Component Separation for Implantable Kidneys Using a Sawtooth Electrode and Negative Dielectrophoresis

Implantable artificial kidneys represent a promising alternative for patients with end-stage renal disease (ESRD), aiming to overcome the limitations of conventional dialysis through the integration of microfluidic and electrokinetic technologies. In this study, we present a sawtooth electrode microfluidic chamber that achieves blood cell separation via negative dielectrophoresis at a record-low operating voltage of 1.4 V, representing a fivefold reduction compared with rectangular electrode designs and supporting potential integration into implantable artificial kidney systems. A microfluidic chip incorporating an asymmetric sawtooth electrode geometry was developed to enhance local electric field gradients while reducing power consumption. Device performance was investigated using COMSOL Multiphysics simulations. Response Surface Methodology (RSM) based on a Box–Behnken design was employed to optimize the number of teeth per unit length (N), sawtooth height (H), and applied voltage (V), while excitation frequency was fixed at 1 MHz and flow velocity was maintained constant at 0.1 µL·min−1. Statistical analysis was conducted using analysis of variance (ANOVA) in Minitab (Version 27; Minitab, LLC, State College, PA, USA, 2024). The optimization model showed strong predictive capability (R2 = 95.8%) and identified applied voltage (59.45% contribution) and sawtooth height (33%) as the dominant factors affecting separation efficiency, with a significant H × V interaction (p = 0.023). Comprehensive voltage-response mapping over the range of 0.8–4.0 V revealed four operational regimes, including a previously unreported high-voltage failure zone above 2.8 V, where electrothermal flow and electroporation degrade performance. Under physiological conductivity conditions, the optimized design maintained a separation efficiency of 78.3% at 1.4 V with a tip temperature rise of only 1.2 °C, while full recovery of performance was achieved at 2.2 V. Cell-specific separation efficiencies reached 97.3% for white blood cells, 95.8% for red blood cells, and 84.7% for platelets, reducing the downstream cellular load by 92.6%. These findings demonstrate that the proposed low-voltage, high-efficiency separation platform has strong potential as a cellular pre-filtration module in implantable artificial kidney systems and other lab-on-chip biomedical devices.

Authors

Institutions

Publication Details

Journal
Applied Sciences
Published
2026-03-13
DOI
https://doi.org/10.3390/app16062785
Primary Topic
Microfluidic and Bio-sensing Technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Low-Voltage Blood Component Separation for Implantable Kidneys Using a Sawtooth Electrode and Negative Dielectrophoresis

Basem Ammar, Hasan Mhd Nazha, Mhd Ayham Darwich, Al-Hasan Ali
Applied Sciences
Microfluidic and Bio-sensing Technologies
article

Low-Voltage Blood Component Separation for Implantable Kidneys Using a Sawtooth Electrode and Negative Dielectrophoresis

Basem Ammar, Hasan Mhd Nazha, Mhd Ayham Darwich, Al-Hasan Ali
article en

Abstract

Implantable artificial kidneys represent a promising alternative for patients with end-stage renal disease (ESRD), aiming to overcome the limitations of conventional dialysis through the integration of microfluidic and electrokinetic technologies. In this study, we present a sawtooth electrode microfluidic chamber that achieves blood cell separation via negative dielectrophoresis at a record-low operating voltage of 1.4 V, representing a fivefold reduction compared with rectangular electrode designs and supporting potential integration into implantable artificial kidney systems. A microfluidic chip incorporating an asymmetric sawtooth electrode geometry was developed to enhance local electric field gradients while reducing power consumption. Device performance was investigated using COMSOL Multiphysics simulations. Response Surface Methodology (RSM) based on a Box–Behnken design was employed to optimize the number of teeth per unit length (N), sawtooth height (H), and applied voltage (V), while excitation frequency was fixed at 1 MHz and flow velocity was maintained constant at 0.1 µL·min−1. Statistical analysis was conducted using analysis of variance (ANOVA) in Minitab (Version 27; Minitab, LLC, State College, PA, USA, 2024). The optimization model showed strong predictive capability (R2 = 95.8%) and identified applied voltage (59.45% contribution) and sawtooth height (33%) as the dominant factors affecting separation efficiency, with a significant H × V interaction (p = 0.023). Comprehensive voltage-response mapping over the range of 0.8–4.0 V revealed four operational regimes, including a previously unreported high-voltage failure zone above 2.8 V, where electrothermal flow and electroporation degrade performance. Under physiological conductivity conditions, the optimized design maintained a separation efficiency of 78.3% at 1.4 V with a tip temperature rise of only 1.2 °C, while full recovery of performance was achieved at 2.2 V. Cell-specific separation efficiencies reached 97.3% for white blood cells, 95.8% for red blood cells, and 84.7% for platelets, reducing the downstream cellular load by 92.6%. These findings demonstrate that the proposed low-voltage, high-efficiency separation platform has strong potential as a cellular pre-filtration module in implantable artificial kidney systems and other lab-on-chip biomedical devices.

Applied SciencesVol. 16(6)
Al Andalus University (SY), Damascus University (SY), Otto-von-Guericke-Universität Magdeburg (DE), Higher Institute for Applied Sciences and Technology (SY)
Affordable and clean energy
Openalex Percentile: Top 23%
Microfluidic and Bio-sensing Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.