A comparison of dose between cone beam computed tomography and conventional computed tomography in broad beam geometry

A unified computed tomography dose index (CTDI) which includes all scattered radiation was used to investigate the difference in absorbed dose to the patient between a computed tomography (CT) and a cone beam CT (CBCT). The unified CTDI, denoted CTDI300,w, was measured using a 100 mm pencil ionization chamber at three positions covering 300 mm at each of the five phantom slots in three CTDI body phantoms. The dose length product of a lumbar spine protocol for the modalities was comparable (CT: 153 mGycm, CBCT: 126 mGycm), though the normalized CTDI300,w for the CT was 4.6 times higher than for the CBCT. The comparison between CTDI300,w and the two dose indexes provided by the systems showed a substantial underestimation for both modalities. The most accurate, but also the most inconvenient method to compare absorbed dose is to use the CTDI300,w. Most important is to always use the same dose index.

Authors

Institutions

Publication Details

Journal
Radiation Protection Dosimetry
Published
2025-09-30
DOI
https://doi.org/10.1093/rpd/ncaf122
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A comparison of dose between cone beam computed tomography and conventional computed tomography in broad beam geometry

Olivia Carlstein, Sara Asplund, Joel Larsson, Zakarias Grönqvist
Radiation Protection Dosimetry
Radiation Dose and Imaging
article

A comparison of dose between cone beam computed tomography and conventional computed tomography in broad beam geometry

Olivia Carlstein, Sara Asplund, Joel Larsson, Zakarias Grönqvist
article en

Abstract

A unified computed tomography dose index (CTDI) which includes all scattered radiation was used to investigate the difference in absorbed dose to the patient between a computed tomography (CT) and a cone beam CT (CBCT). The unified CTDI, denoted CTDI300,w, was measured using a 100 mm pencil ionization chamber at three positions covering 300 mm at each of the five phantom slots in three CTDI body phantoms. The dose length product of a lumbar spine protocol for the modalities was comparable (CT: 153 mGycm, CBCT: 126 mGycm), though the normalized CTDI300,w for the CT was 4.6 times higher than for the CBCT. The comparison between CTDI300,w and the two dose indexes provided by the systems showed a substantial underestimation for both modalities. The most accurate, but also the most inconvenient method to compare absorbed dose is to use the CTDI300,w. Most important is to always use the same dose index.

Radiation Protection DosimetryVol. 202(3–4)
Sahlgrenska University Hospital (SE), Region Västra Götaland (SE), NU Hospital Group (SE), University of Gothenburg (SE)
Openalex Percentile: Top 25%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A comparison of dose between cone beam computed tomography and conventional computed tomography in broad beam geometry — Olivia Carlstein, Sara Asplund, et al. · Radiation Protection Dosimetry (2025) | TGRS Research Map | TGRS