A methodology for assessing the effect of localizer orientation on the consistency of tube current modulation in CT

Localizer orientation can significantly impact the automatic tube-current modulation (ATCM) in computed tomography (CT). The purpose of this study was to introduce a methodology for assessing ATCM consistency and to compare the outcomes in image noise and radiation dose between a single anterior-posterior (AP) localizer and a combination of lateral and AP (LAT + AP) localizers. A total of 299 prospective patients referred for a routine CT thorax examination were randomly assigned to a single AP or LAT + AP localizer. Measurements of image noise and patient size were performed in the thorax and liver, and the corresponding effective milliampere-second (mAseff) at these positions were collected. Performance analysis was conducted using the normalized root mean squared error (nRMSE) of mAseff and image noise relative to patient size. A smaller nRMSE indicates more consistent inter-patient variance in image noise and mAseff for a given size. The LAT + AP group had a statistically significant lower nRMSE for mAseff and image noise in the thorax region, and for mAseff (but not image noise) in the liver region. A post-hoc finding indicated that a subgroup of female patients with laterally protruding breasts was overexposed (+57% radiation dose) in the thorax region when using the AP localizer. Using LAT + AP localizer resulted in more consistent image noise levels and radiation doses compared to single AP localizer. The effect was small, except for the subgroup of females, where radiation doses were higher in the AP group. The findings are specific to the CT system used in this study and may not be generalizable to other CT models or software versions. Nonetheless, the proposed methodology provides a valuable approach for evaluating ATCM consistency.

Authors

Institutions

Publication Details

Journal
Radiation Protection Dosimetry
Published
2025-12-19
DOI
https://doi.org/10.1093/rpd/ncaf190
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A methodology for assessing the effect of localizer orientation on the consistency of tube current modulation in CT

Marcus Söderberg, Daniel E. Thor, Torkel B. Brismar, Gavin G. Poludniowski
Radiation Protection Dosimetry
Radiation Dose and Imaging
article

A methodology for assessing the effect of localizer orientation on the consistency of tube current modulation in CT

Marcus Söderberg, Daniel E. Thor, Torkel B. Brismar, Gavin G. Poludniowski
article en

Abstract

Localizer orientation can significantly impact the automatic tube-current modulation (ATCM) in computed tomography (CT). The purpose of this study was to introduce a methodology for assessing ATCM consistency and to compare the outcomes in image noise and radiation dose between a single anterior-posterior (AP) localizer and a combination of lateral and AP (LAT + AP) localizers. A total of 299 prospective patients referred for a routine CT thorax examination were randomly assigned to a single AP or LAT + AP localizer. Measurements of image noise and patient size were performed in the thorax and liver, and the corresponding effective milliampere-second (mAseff) at these positions were collected. Performance analysis was conducted using the normalized root mean squared error (nRMSE) of mAseff and image noise relative to patient size. A smaller nRMSE indicates more consistent inter-patient variance in image noise and mAseff for a given size. The LAT + AP group had a statistically significant lower nRMSE for mAseff and image noise in the thorax region, and for mAseff (but not image noise) in the liver region. A post-hoc finding indicated that a subgroup of female patients with laterally protruding breasts was overexposed (+57% radiation dose) in the thorax region when using the AP localizer. Using LAT + AP localizer resulted in more consistent image noise levels and radiation doses compared to single AP localizer. The effect was small, except for the subgroup of females, where radiation doses were higher in the AP group. The findings are specific to the CT system used in this study and may not be generalizable to other CT models or software versions. Nonetheless, the proposed methodology provides a valuable approach for evaluating ATCM consistency.

Radiation Protection DosimetryVol. 202(3–4)
Karolinska University Hospital (SE), Karolinska Institutet (SE), Skåne University Hospital (SE)
Openalex Percentile: Top 25%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.