Accurate computed tomography (CT) numbers derived from dual-energy CT (DECT) have numerous applications, including optimising the precision of radiation treatment planning. The dual-energy iterative reconstruction algorithm (DIRA) utilizes material decomposition in the iterative loop to generate monoenergetic images free of beam-hardening artifacts. In simulations, the reconstructed CT numbers closely matched reference tabulated values. This study evaluates the feasibility of applying DIRA to experimental data. Transitioning from simulations required several adaptations: (i) removal of the patient table, (ii) modeling of the bow-tie filter, and (iii) application of an inverse water beam-hardening correction. Axial scans of a cylindrical polymethyl methacrylate phantom containing four rod inserts of distinct materials were acquired using a clinical Siemens SOMATOM Force scanner. By the sixth iteration, DIRA yielded CT numbers that closely matched tabulated values. The algorithm also demonstrated strong performance on an anthropomorphic head phantom (Computerized Imaging Reference Systems, Inc. (CIRS) model 731-HN) with known material composition.

Authors

Institutions

Publication Details

Journal
Radiation Protection Dosimetry
Published
2025-10-31
DOI
https://doi.org/10.1093/rpd/ncaf140
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluating the dual-energy iterative reconstruction algorithm (DIRA) for accurate CT number determination in DECT imaging

Åsa Carlsson Tedgren, Alexandr Malušek, Michael Sandborg, Maria Magnusson
Radiation Protection Dosimetry
Advanced X-ray and CT Imaging
article

Evaluating the dual-energy iterative reconstruction algorithm (DIRA) for accurate CT number determination in DECT imaging

Åsa Carlsson Tedgren, Alexandr Malušek, Michael Sandborg, Maria Magnusson
article en

Abstract

Accurate computed tomography (CT) numbers derived from dual-energy CT (DECT) have numerous applications, including optimising the precision of radiation treatment planning. The dual-energy iterative reconstruction algorithm (DIRA) utilizes material decomposition in the iterative loop to generate monoenergetic images free of beam-hardening artifacts. In simulations, the reconstructed CT numbers closely matched reference tabulated values. This study evaluates the feasibility of applying DIRA to experimental data. Transitioning from simulations required several adaptations: (i) removal of the patient table, (ii) modeling of the bow-tie filter, and (iii) application of an inverse water beam-hardening correction. Axial scans of a cylindrical polymethyl methacrylate phantom containing four rod inserts of distinct materials were acquired using a clinical Siemens SOMATOM Force scanner. By the sixth iteration, DIRA yielded CT numbers that closely matched tabulated values. The algorithm also demonstrated strong performance on an anthropomorphic head phantom (Computerized Imaging Reference Systems, Inc. (CIRS) model 731-HN) with known material composition.

Radiation Protection DosimetryVol. 202(3–4)
Linköping University (SE), Karolinska University Hospital (SE), Linköping University Hospital (SE)
Openalex Percentile: Top 37%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.