Quantitative Ultrasound Texture Analysis of Breast Tumor Responses to Chemotherapy: Comparison of a Cart-Based and a Wireless Ultrasound Scanner

This study assessed the level of agreement between quantitative ultrasound (QUS) feature estimates derived from ultrasound images of breast tumors in women with locally advanced breast cancer (LABC) produced using a cart-based and a handheld ultrasound system. Thirty LABC patients receiving neoadjuvant chemotherapy were imaged at two separate times: a pre-treatment 'baseline' time point, and four weeks after the start of chemotherapy. Three sets of QUS features were produced using the reference phantom technique, one for each imaging time and a third set calculated by taking the differences in feature estimates between times. Cross-system statistical testing using the Wilcoxon signed-rank test was performed for each feature set to assess the level of feature estimate agreement between ultrasound systems. The Bland-Altman method was employed to graphically assess feature sets for systematic skew. The range of <i>p</i>-values was 4.50 × 10<sup>-11</sup> to 0.277 for the baseline features, 2.77 × 10<sup>-5</sup> to 0.865 for the week 4 features, and 2.03 × 10<sup>-9</sup> to 1 for the feature differences. For the feature differences, all five of the primary QUS features (MBF, SS, SI, ASD, AAC) were found to be in agreement between the two scanner types at the 5% confidence level. For the baseline feature set and week 4 feature set, 0 out of 5 and 3 out of 5 of the primary features were found to be in agreement, respectively. Of the 20 QUS texture features examined, the number and proportion of the total for each feature set which were found to have statistically significant similarity in their sample medians at the 5% confidence level were as follows: 2 out of 20 (10%) for the baseline features; 17 out of 20 (85%) for the week 4 features; and 12 out of 20 (60%) for the feature differences. The specific texture features found to be in agreement varied between QUS-specific feature sets. Overall, a moderate level of agreement between sets of feature differences produced using the two systems was demonstrated.

Authors

Institutions

Publication Details

Journal
Journal of Imaging
Published
2026-03-13
DOI
https://doi.org/10.3390/jimaging12030129
Primary Topic
MRI in cancer diagnosis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quantitative Ultrasound Texture Analysis of Breast Tumor Responses to Chemotherapy: Comparison of a Cart-Based and a Wireless Ultrasound Scanner

Michael L. Oelze, Joyce Jing Yee Yip, Maria Lourdes Anzola Pena, Laurentius O. Osapoetra et al.
Journal of Imaging
MRI in cancer diagnosis
article

Quantitative Ultrasound Texture Analysis of Breast Tumor Responses to Chemotherapy: Comparison of a Cart-Based and a Wireless Ultrasound Scanner

Michael L. Oelze, Joyce Jing Yee Yip, Maria Lourdes Anzola Pena, Laurentius O. Osapoetra, David Alberico, Lakshmanan Sannachi, Frances Wright, Gregory J. Czarnota, Sonal Gandhi
article en

Abstract

This study assessed the level of agreement between quantitative ultrasound (QUS) feature estimates derived from ultrasound images of breast tumors in women with locally advanced breast cancer (LABC) produced using a cart-based and a handheld ultrasound system. Thirty LABC patients receiving neoadjuvant chemotherapy were imaged at two separate times: a pre-treatment 'baseline' time point, and four weeks after the start of chemotherapy. Three sets of QUS features were produced using the reference phantom technique, one for each imaging time and a third set calculated by taking the differences in feature estimates between times. Cross-system statistical testing using the Wilcoxon signed-rank test was performed for each feature set to assess the level of feature estimate agreement between ultrasound systems. The Bland-Altman method was employed to graphically assess feature sets for systematic skew. The range of <i>p</i>-values was 4.50 × 10<sup>-11</sup> to 0.277 for the baseline features, 2.77 × 10<sup>-5</sup> to 0.865 for the week 4 features, and 2.03 × 10<sup>-9</sup> to 1 for the feature differences. For the feature differences, all five of the primary QUS features (MBF, SS, SI, ASD, AAC) were found to be in agreement between the two scanner types at the 5% confidence level. For the baseline feature set and week 4 feature set, 0 out of 5 and 3 out of 5 of the primary features were found to be in agreement, respectively. Of the 20 QUS texture features examined, the number and proportion of the total for each feature set which were found to have statistically significant similarity in their sample medians at the 5% confidence level were as follows: 2 out of 20 (10%) for the baseline features; 17 out of 20 (85%) for the week 4 features; and 12 out of 20 (60%) for the feature differences. The specific texture features found to be in agreement varied between QUS-specific feature sets. Overall, a moderate level of agreement between sets of feature differences produced using the two systems was demonstrated.

Journal of ImagingVol. 12(3)
Sunnybrook Health Science Centre (CA), University of Illinois Urbana-Champaign (US), University of Toronto (CA), Health Sciences Centre (CA), Sunnybrook Research Institute
Good health and well-being
Openalex Percentile: Top 24%
MRI in cancer diagnosis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.