Bench Evaluation of Gas Duration in a Disaster for Two Frequently Used Portable Ventilators for Neonates

Background: Blended medical gas is required for premature infants in the NICU, a high resource-utilization area during a disaster. Portable ventilators are powered by battery or plugged into generator- supported emergency electrical outlets. Oxygen is bled in, or attached to a 50 PSI gas source, while air is provided by compressed gas cylinder, a turbine driven source, or compressor. Portable ventilators are often needed when a disaster resulting in a power outage or gas failure occurs. The purpose of this bench study is to compare the duration of gas consumption on two frequently used portable ventilators for neonates. Methods: A pNeuton mini pneumatic portable ventilator (Airon, Melbourne, FL) was attached to a pNeuton Mini Patient Circuit (Airon). A Hamilton T1 ventilator (Hamilton Medical, Reno, NV) was connected to a RT 265 neonatal ventilator circuit (Fisher & Paykel, Irvine, CA). The circuit was attached to an Infant Ingmar Neo Lung (Ingmar, Pittsburgh, PA) for the individual gas duration tests at differing oxygenation percentages on both ventilators. We compared the duration of oxygen gas at a FIO2 of 21%, 40%, 60% and 100% when attached to 2200 PSI e-cylinder oxygen. The settings used for both ventilators tested were pressure control PIP 25, PEEP +5, rate 30, and inspiratory time 0.5 s. The pNeuton was set to a constant flow of 6 L/min. The Hamilton T1 flow was calculated by the ventilator for the given settings. Duration of gas was determined as compressed oxygen gas from an E cylinder transitioned from full to empty cylinder at each of the specified oxygen percentages. Results: The results of oxygen gas duration in minutes were displayed in the accompanying table. The Hamilton T1 ventilator consumed oxygen three times less than the pNeuton Mini ventilator. Conclusions: At neonatal settings, an electrically powered and turbine-driven powered portable ventilators are good options for mechanical ventilation when a blended gas is needed in the presence of a gas failure. The Hamilton T1’s turbine driven mechanism is gas conserving and preferred over pneumatic driven ventilators as long as a power source is available. However, understanding your gas and ventilator resources and choosing the correct ventilator to support these emergencies is critical.Oxygen Duration In MinutesFiO2Hamilton T1pNeuton Mini21%Internal compressor no tank needed11940%89123560%522148100%27096

Authors

Institutions

Publication Details

Journal
Respiratory Care
Published
2024-10-01
DOI
https://doi.org/10.4187/respcare.20244135421
Primary Topic
Disaster Response and Management
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Bench Evaluation of Gas Duration in a Disaster for Two Frequently Used Portable Ventilators for Neonates

Chad Earl Weagraff, Kathleen Deakins
Respiratory Care
Disaster Response and Management
article

Bench Evaluation of Gas Duration in a Disaster for Two Frequently Used Portable Ventilators for Neonates

Chad Earl Weagraff, Kathleen Deakins
article en

Abstract

Background: Blended medical gas is required for premature infants in the NICU, a high resource-utilization area during a disaster. Portable ventilators are powered by battery or plugged into generator- supported emergency electrical outlets. Oxygen is bled in, or attached to a 50 PSI gas source, while air is provided by compressed gas cylinder, a turbine driven source, or compressor. Portable ventilators are often needed when a disaster resulting in a power outage or gas failure occurs. The purpose of this bench study is to compare the duration of gas consumption on two frequently used portable ventilators for neonates. Methods: A pNeuton mini pneumatic portable ventilator (Airon, Melbourne, FL) was attached to a pNeuton Mini Patient Circuit (Airon). A Hamilton T1 ventilator (Hamilton Medical, Reno, NV) was connected to a RT 265 neonatal ventilator circuit (Fisher & Paykel, Irvine, CA). The circuit was attached to an Infant Ingmar Neo Lung (Ingmar, Pittsburgh, PA) for the individual gas duration tests at differing oxygenation percentages on both ventilators. We compared the duration of oxygen gas at a FIO2 of 21%, 40%, 60% and 100% when attached to 2200 PSI e-cylinder oxygen. The settings used for both ventilators tested were pressure control PIP 25, PEEP +5, rate 30, and inspiratory time 0.5 s. The pNeuton was set to a constant flow of 6 L/min. The Hamilton T1 flow was calculated by the ventilator for the given settings. Duration of gas was determined as compressed oxygen gas from an E cylinder transitioned from full to empty cylinder at each of the specified oxygen percentages. Results: The results of oxygen gas duration in minutes were displayed in the accompanying table. The Hamilton T1 ventilator consumed oxygen three times less than the pNeuton Mini ventilator. Conclusions: At neonatal settings, an electrically powered and turbine-driven powered portable ventilators are good options for mechanical ventilation when a blended gas is needed in the presence of a gas failure. The Hamilton T1’s turbine driven mechanism is gas conserving and preferred over pneumatic driven ventilators as long as a power source is available. However, understanding your gas and ventilator resources and choosing the correct ventilator to support these emergencies is critical.Oxygen Duration In MinutesFiO2Hamilton T1pNeuton Mini21%Internal compressor no tank needed11940%89123560%522148100%27096

Respiratory CareVol. 69(10_suppl)
Rainbow Babies & Children's Hospital (US)
Climate action
Openalex Percentile: Top 36%
Disaster Response and Management
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.