Effect of clinical‐crown height space on extra‐narrow dental implant (2.9 mm) fracture load and failure mode
Within the limitations of this in vitro study, CHS significantly influenced the failure load of ENDI (2.9 mm)-supported cement-retained zirconia crowns, with increased CHS associated with reduced load-bearing capacity.